How Cyclic Loading Affects the Migration of Radio-Opaque Markers Attached to Tendon Grafts Using a New Method: A Study Using Roentgen Stereophotogrammetric Analysis (RSA)

机译:循环载荷如何使用一种新方法影响附着在肌腱移植物上的不透射线标志物的迁移:一项使用伦琴立体摄影测量分析(RSA)的研究

获取原文
获取原文并翻译 | 示例

摘要

An increase in anterior laxity following reconstruction of the anterior cruciate ligament (ACL) can result from lengthening of the graft construct in either the regions of fixation and/or the region of the graft substance between the fixations. RSA could be a useful technique to determine lengthening in these regions if a method can be devised for attaching radio-opaque markers to soft tissue grafts so that marker migration from repeated loading of the graft is limited. Therefore, the objectives of this study were 1) to develop a method for attaching radio-opaque markers to an ACL graft that limits marker migration within the graft, 2) to characterize the error of an RSA system used to study migration, and 3) to determine the maximum amount of migration and the time when it occurs during cyclic loading of ACL grafts. Tendon markers were constructed from a 0.8-mm tantalum ball and a stainless steel suture. Ten double-looped tendon grafts were passed through tibial tunnels drilled in bovine tibias and fixed with a tibial fixation device. Two tendon markers were sewn to one tendon bundle of each graft and the grafts were cyclically loaded for 225,000 cycles from 20 N to 170 N. At specified intervals, simultaneous radiographs were obtained of the tendon markers and a radiographic standard of known length. The bias and imprecision in measuring the length of the radiographic standard were 0.0 and 0.046 mm respectively. Marker migration was computed as the change in distance between the two tendon markers along the axis of the tibial tunnel. Marker migration was greatest after 225,000 cycles with a root mean square (RMS) value of less than 0.2 mm. Because the RMS value indicates the error introduced into measurements of lengthening and because this error is small, the method described for attaching markers to an ACL graft has the potential to be useful for determining lengthening of ACL graft constructs in in vivo studies in humans.
机译:重建前交叉韧带(ACL)后,前松弛度的增加可能是由于固定部位和/或固定部位之间的移植物区域中移植物构造的延长。如果可以设计一种将不透射线的标记物附着到软组织移植物上的方法,则RSA可能是确定这些区域延长的有用技术,这样就可以限制来自反复加载移植物的标记物迁移。因此,本研究的目的是:1)开发一种将不透射线的标记物附着到ACL移植物上的方法,以限制标记物在移植物中的迁移; 2)表征用于研究迁移性的RSA系统的错误; 3)确定最大的迁移量以及在ACL移植物循环加载期间发生迁移的时间。肌腱标记物是由0.8毫米的钽球和不锈钢缝线制成的。十个双环肌腱移植物穿过在牛胫骨中钻出的胫骨隧道,并用胫骨固定装置固定。将两个肌腱标记物缝合到每个移植物的一个肌腱束上,并从20 N到170 N循环加载移植物225,000个周期。在指定的时间间隔内,同时获得肌腱标记物和已知长度的X线照相标准照。测量射线照相标准长度的偏差和不精确度分别为0.0和0.046 mm。标记物的移动被计算为两个腱标记物之间沿胫骨隧道轴线的距离变化。 225,000次循环后,标记迁移最大,均方根(RMS)值小于0.2 mm。由于RMS值指示引入延长测量的误差,并且由于该误差很小,因此描述的将标记物附着到ACL移植物上的方法具有潜在的潜力,可用于确定人体体内研究中ACL移植物构建物的长度。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号