首页> 外文期刊>The Journal of Bone and Joint Surgery. American Volume >A comparison of MRI and CT imaging clarity of titanium alloy and titanium alloy with cobalt-chromium-alloy pedicle screw and rod implants in the lumbar spine
【24h】

A comparison of MRI and CT imaging clarity of titanium alloy and titanium alloy with cobalt-chromium-alloy pedicle screw and rod implants in the lumbar spine

机译:钛合金和钛合金与钴铬合金椎弓根螺钉和棒状植入物在腰椎中的MRI和CT成像清晰度比较

获取原文
获取原文并翻译 | 示例
           

摘要

Background: Magnetic resonance imaging (MRI) and computed tomography (CT) imaging are important postoperative diagnostic and evaluation tools, particularly in patients who have undergone spinal fusions. Advancements in materials and imaging techniques have lessened artifact and improved overall imaging results. Systems that combine titanium alloy and cobalt-chromium components have been introduced to reduce implant profile while maintaining strength. The objective of this study was to determine if there were any differences in the clarity of imaging between two types of implant materials in a lumbar spine construct model. Methods: One of two lumbar spine stabilization implant systems, titanium alloy (titanium) or titanium alloy with cobaltchromiumalloy (titanium-cobalt), was placed to simulate a four-level fusion construct in two human cadaveric spine segments, followed byMRI andCT imaging. The implant systemswere then removed fromeach cadaver and implanted in the other cadaver. Nine physician graders from three subspecialties scored the images using a 5-point scale,with higher imaging scores indicating greater clarity of the region of interest. Physician-rated scores were compared across systems and between physician groups. Results: There were no significant differences in the overall mean total scores on the basis of construct material. Overall mean scores were 18.16 for titanium and 17.45 for titanium-cobalt (p = 0.275). Among images of the titanium-cobalt constructs, no significant differences in mean scores were found between specimens with use of MRI (p = 0.883) or with use of CT only (p = 0.274). Among images of the titanium system, a slightly significant difference was found between specimens with use of MRI (p = 0.044) but not with CT imaging (p = 0.837). Conclusions: Overall image clarity scores were not significantly different between titanium and titanium-cobalt implant systems in the lumbar spine. Observation of pertinent anatomy in the regions of interest was not degraded by the presence of either system. Clinical Relevance: Postoperative imaging of spinal implants does not appear to be affected by spinal implant systems of titanium-cobalt or titanium alone.
机译:背景:磁共振成像(MRI)和计算机断层扫描(CT)成像是重要的术后诊断和评估工具,尤其是在经历过脊柱融合术的患者中。材料和成像技术的进步已减少了伪影并改善了整体成像结果。已经引入了将钛合金和钴铬成分相结合的系统,以减少植入物的轮廓,同时保持强度。这项研究的目的是确定腰椎构造模型中两种类型的植入材料之间的成像清晰度是否存在差异。方法:将两个钛合金或钛钴钴合金(钛钴)两种腰椎稳定植入系统之一放置在两个人体尸体脊椎节段中模拟四级融合结构,然后进行MRI和CT成像。然后从每个尸体上取下植入系统,然后将其植入另一个尸体中。来自三个亚专业的9位医师评分者使用5分制对图像进行评分,成像得分越高,表明感兴趣区域的清晰度越高。在整个系统之间以及医师组之间比较医师评分。结果:根据构建材料,总体平均总分没有显着差异。钛的总体平均得分为18.16,钛-钴的总体平均得分为17.45(p = 0.275)。在钛钴构造的图像中,使用MRI(p = 0.883)或仅使用CT(p = 0.274)在标本之间没有发现平均得分的显着差异。在钛系统的图像中,使用MRI(p = 0.044)而在标本之间发现了略有显着差异,而使用CT成像(p = 0.837)却没有。结论:腰椎中钛和钛-钴植入系统之间的总体图像清晰度得分没有显着差异。任何一个系统的存在都不会降低对感兴趣区域相关解剖结构的观察。临床意义:脊柱植入物的术后成像似乎不受钛钴或钛的脊柱植入系统的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号