BACKGROUND: Metal-on-metal bearings, as used in total hip arthroplasty prostheses that have a large-diameter femoral head, were proposed as an option for treating young and active patients with degenerative hip disorders. Despite the theoretical improved performance of large metal articulations with regard to wear, metal ion levels produced by total hip arthroplasty prostheses in which a large-diameter femoral head is used have not been thoroughly evaluated. METHODS: From 173 eligible patients, 144 patients were allocated to undergo a unilateral total hip arthroplasty with use of metal-on-metal components and a large-diameter femoral head. The purpose of this study was to compare the amount of metal ion release (chromium, cobalt, and titanium) from four different types of prostheses from four different implant manufacturers (Biomet, DePuy, Smith & Nephew, and Zimmer). RESULTS: For cobalt ion levels, a significant difference was found between the different types of total hip arthroplasty prostheses with a large-diameter femoral head at three, six, twelve, and twenty-four months, but only in male patients or patients with a femoral head size of 50 mm or greater. The highest mean cobalt levels at all follow-up periods were observed with the Zimmer implant and the lowest with the Biomet implant. Titanium ion levels were highest in the Zimmer group at all follow-up periods, whereas no difference between groups was found for chromium. At the most recent follow-up, one hip was revised after the development of an adverse local tissue reaction. Of concern was the observation at the time of revision surgery of black metallic deposits inside the adapter sleeve and on the prosthetic femoral neck. CONCLUSIONS: This investigation revealed that metal ion release differs greatly between various total hip arthroplasty implants with a large-diameter femoral head. The sources of metal ion production are numerous, but it is hypothesized that wear and corrosion at the junction between the adapter sleeve and the femoral stem in some patients may be responsible for the elevated cobalt ion levels that were found in the Zimmer group. On the other hand, an adapter sleeve made of titanium, such as the one used with the Biomet large-diameter-head total hip arthroplasty, is an unlikely contributor to the release of cobalt ions. Current technology or design of some total hip arthroplasty systems that make use of a large-diameter femoral head may not yet allow the use of modular large heads with a metal-on-metal articulation, especially in young, active male patients whose activities generate high loads at the hip joint. Further research is needed to better understand the favorable design characteristics of modular junctions in metal-on-metal total hip arthroplasty implants that make use of large-diameter femoral heads.
机译:背景:金属对金属轴承,用于具有大直径股骨头的全髋关节置换术假体,已被提议作为治疗退行性髋关节疾病的年轻和活跃患者的一种选择。尽管从理论上讲,大型金属关节在磨损方面的性能有所改善,但尚未对通过使用大直径股骨头的全髋关节置换假体产生的金属离子水平进行全面评估。方法:从173名符合条件的患者中,将144例患者分配为使用金属对金属部件和大直径股骨头进行单侧全髋关节置换术。这项研究的目的是比较来自四个不同植入物制造商(Biomet,DePuy,Smith&Nephew和Zimmer)的四种不同类型假体的金属离子释放量(铬,钴和钛)。结果:对于钴离子水平,在大直径股骨头处分别在3、6、12和24个月时,在不同类型的全髋关节置换术假体之间发现了显着差异,但仅在男性患者或患有假体的患者中。股骨头大小为50毫米或更大。 Zimmer植入物在所有随访期间均观察到最高平均钴水平,而Biomet植入物则观察到最低水平。在所有随访期间,齐默组的钛离子水平最高,而铬之间在两组之间没有差异。在最近的随访中,在出现不良的局部组织反应后,对一只髋进行了翻修。值得关注的是,在进行翻修手术时,在紧定套内和假体股骨颈上发现了黑色金属沉积物。结论:这项研究表明,在具有大直径股骨头的各种全髋关节置换植入物之间,金属离子的释放差异很大。金属离子的产生来源很多,但据推测,在某些患者中,紧定套和股骨柄之间的连接处的磨损和腐蚀可能是Zimmer组中钴离子水平升高的原因。另一方面,钛制的紧定套(例如与Biomet大直径头全髋关节置换术一起使用的紧定套)不太可能促成钴离子的释放。利用大直径股骨头的某些全髋关节置换系统的当前技术或设计可能尚未允许使用具有金属对金属铰接的模块化大头,特别是在活跃且活跃的男性患者中髋关节负荷。需要进一步的研究,以更好地了解使用大直径股骨头的金属对金属全髋关节置换植入物中模块化连接的有利设计特性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号