首页> 外文期刊>The journal of asthma >Measurement of exhaled nitric oxide in a general population sample: a comparison of the Medisoft HypAir FE(NO) and Aerocrine NIOX analyzers.
【24h】

Measurement of exhaled nitric oxide in a general population sample: a comparison of the Medisoft HypAir FE(NO) and Aerocrine NIOX analyzers.

机译:普通人群样本中呼出气一氧化氮的测量:Medisoft HypAir FE(NO)和Aerocrine NIOX分析仪的比较。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND AND OBJECTIVE: Measuring the fraction of nitric oxide in exhaled breath (FE(NO)) is increasingly utilized to assess airway inflammation in asthma. The primary aim of this study was to compare exhaled nitric oxide measurements obtained using two devices from different manufacturers, that is, the recently marketed portable and electrochemical-based Medisoft HypAir FE(NO) and the well-established chemiluminescence-based Aerocrine NIOX analyzer, in an unselected population. METHODS: FE(NO) measurements were conducted in 106 subjects (86 healthy; 20 asthmatic; 56.6% atopic). Atopy and health status were assessed by skin prick tests and questionnaire, respectively. RESULTS: The two instruments showed strong correlation over a wide range of FE(NO) measurements (8-261.3 ppb with the HypAir, 5.6-156.8 ppb with the NIOX; r = 0.98; p < .0001). This correlation was observed in the population as a whole, as well as in healthy non-atopics, healthy atopics, and atopic asthmatics when considered separately. The measurements on the HypAir FE(NO) were consistently 1.6 times (95% CI 1.11-2.05) higher than those obtained with the NIOX. CONCLUSIONS: FE(NO) measurements obtained with the HypAir FE(NO) correlated well with the NIOX, but were approximately 1.6 times higher. Therefore, a conversion factor is required if results are to be compared with the NIOX instrument.
机译:背景与目的:测量呼出气中一氧化氮(FE(NO))的比例越来越多地用于评估哮喘的气道炎症。这项研究的主要目的是比较使用不同制造商的两种设备获得的呼出气一氧化氮的测量结果,即最近出售的基于电化学的便携式Medicom HypAir FE(NO)和成熟的基于化学发光的Aerocrine NIOX分析仪,在未选定的人群中。方法:对106名受试者(86名健康; 20名哮喘; 56.6%特应性)进行了FE(NO)测量。分别通过皮肤点刺试验和问卷调查评估了特应性和健康状况。结果:两种仪器在广泛的FE(NO)测量中均显示出很强的相关性(HypAir使用8-261.3 ppb,NIOX使用5.6-156.8 ppb; r = 0.98; p <.0001)。当单独考虑时,在整个人群以及健康的非过敏性疾病,健康过敏性疾病和过敏性哮喘患者中均观察到这种相关性。 HypAir FE(NO)的测量值始终比使用NIOX获得的测量值高1.6倍(95%CI 1.11-2.05)。结论:用HypAir FE(NO)获得的FE(NO)测量值与NIOX相关性很好,但大约高出1.6倍。因此,如果要将结果与NIOX仪器进行比较,则需要转换系数。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号