BACKGROUND: Use of automatic external defibrillators (AEDs) in children aged <8 years is not recommended. The purpose of this study was to develop an ECG database of shockable and nonshockable rhythms from a broad age range of pediatric patients and to test the accuracy of the Agilent Heartstream FR2 Patient Analysis System for sensitivity and specificity. METHODS AND RESULTS: Children aged
机译:背景:不建议在8岁以下的儿童中使用自动体外除颤器(AED)。这项研究的目的是开发各种年龄段儿科患者的可电击和不可电击心律的ECG数据库,并测试安捷伦Heartstream FR2患者分析系统的敏感性和特异性。方法和结果:研究了≥12岁的患有心律不齐或有患心律不齐风险的儿童。该数据库使用两个来源:通过改良的AED前瞻性记录了节律的儿童,以及在纸上捕获心律不齐并数字化以进行后续分析的儿童。将节奏分为5秒,由3位审稿人进行分类,然后通过AED分析算法进行评估。分析了191名年龄在1天至12岁(中位数3.0岁)的儿童(81名女性和110名男性)中的696个五秒节奏带。对于不可电击的节律有100%的特异性。心室纤颤的敏感性为96%。结论:在所有年龄组中,对于室颤和非电击性节律,AED节律分析的敏感性和特异性均极佳。高特异性和敏感性表明,发生不适当电击的风险非常低,并且AED可以正确识别出令人震惊的节律,从而使该算法对儿童既安全又有效。