The RAMPReg. Anthrax Test Cartridge for detecting Bacillus anthracis was validated for use in the field for detection of B. anthracis spores in visible powder residues on 7 nonporous environmental surfaces. Six teams of trained first responders and civil support personnel in Class C personal protective equipment sampled visible powder residues on plastic, stainless steel, ceramic tile, wood, rubber, sealed concrete, and food-grade painted wood and analyzed the samples on the RAMP Anthrax Test System. The accuracy for each surface was at least 97% and the overall average was 98.8%. The overall average false-positive rate was 1.79% and false-negative rate was 1.07% for all surfaces. There were no significant differences between surfaces or between spore levels..
机译:RAMPReg。经验证,用于检测炭疽芽孢杆菌的炭疽测试盒可用于检测7个无孔环境表面上可见粉末残留物中的炭疽芽孢杆菌孢子。六组经过培训的急救人员和民事支持人员在C级个人防护设备中采样了塑料,不锈钢,瓷砖,木材,橡胶,密封混凝土和食品级涂漆木材上的可见粉末残留物,并在RAMP炭疽上分析了样品测试系统。每个表面的精度至少为97%,总体平均值为98.8%。所有表面的总体平均假阳性率为1.79%,假阴性率为1.07%。表面之间或孢子水平之间没有显着差异。