首页> 外文期刊>Journal of analytical chemistry >Elimination of Water Interference in Pulsed Glow Discharge Time-of-Flight Mass Spectrometry
【24h】

Elimination of Water Interference in Pulsed Glow Discharge Time-of-Flight Mass Spectrometry

机译:消除脉冲辉光放电飞行时间质谱中的水干扰

获取原文
获取原文并翻译 | 示例
           

摘要

The presence of water in the discharge cell is a serious problem in glow discharge mass spectrometry. Even very small quantities of water can make considerable changes in the composition and electrical parameters of the plasma, which lead to a decrease in the signal intensity and the appearance of various cluster components. This results in a very complicated mass spectrum and significantly deteriorates the analytical performance of the method. Different approaches to solving the this problem are discussed in the paper. A multiple position interface that allows analyzing 6–10 samples without decapsulation of the discharge cell is presented in this work. It is also shown that the use of a tantalum auxiliary cathode ensures a several-order depression of the interfering components (OH~+, OH_2~+,OH_3~+, ~(12)C~1H_2~+,~(16)O~+,~(12)C~1H_3~+)because of its getter behavior. The simultaneous application of all proposed approaches ensures solving the problem of interferences in the combined hollow cathode with pulsed glow discharge to the great extent.
机译:在辉光放电质谱法中,放电室中水的存在是一个严重的问题。甚至非常少量的水也会使等离子体的组成和电参数发生重大变化,从而导致信号强度的下降和各种簇成分的出现。这导致非常复杂的质谱图,并且大大降低了该方法的分析性能。本文讨论了解决此问题的不同方法。这项工作提出了一个多位置接口,该接口允许分析6-10个样品而无需拆开放电室。还显示出使用钽辅助阴极可确保干扰组分(OH〜+,OH_2〜+,OH_3〜+,〜(12)C〜1H_2〜+,〜(16)O 〜+,〜(12)C〜1H_3〜+),因为它具有吸气行为。所有提出的方法的同时应用确保了在最大程度上解决带有脉冲辉光放电的组合空心阴极中的干扰问题。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号