This paper compares the potential forensic application of two sensitive and rapid procedures (liquid chromatography-mass spectrometry and liquid chromatography-ion trap mass spectrometry) for the detection and quantitation of 2-oxo-3-hydroxy lysergic acid diethylamide (O-H-LSD) a major LSD metabolite. O-H-LSD calibration curves for both procedures were linear over the concentration range 0-8,000 pg/mL with correlation coefficients (r2) greater than 0.99. The observed limit of detection (LOD) and limit of quantitation (LOQ) for O-H-LSD in both procedures was 400 pg/mL. Sixty-eight human urine specimens that had previously been found to contain LSD by gas chromatography-mass spectrometry were reanalyzed by both procedures for LSD and O-H-LSD. These specimens contained a mean concentration of O-H-LSD approximately 16 times higher than the LSD concentration. Because both LC methods produce similar results, either procedure can be readily adapted to O-H-LSD analysis for use in high-volume drug-testing laboratories. In addition, the possibility of significantly increasing the LSD detection time window by targeting this major LSD metabolite for analysis may influence other drug-free workplace programs to test for LSD.
机译:本文比较了两种灵敏而快速的程序(液相色谱-质谱和液相色谱-离子阱质谱)在检测和定量2-氧代-3-羟基麦角酸二乙酰胺(OH-LSD)中的潜在取证应用。主要的LSD代谢产物。两种方法的O-H-LSD校准曲线在0-8,000 pg / mL的浓度范围内都是线性的,相关系数(r2)大于0.99。两种方法中O-H-LSD的观察到的检出限(LOD)和定量限(LOQ)为400 pg / mL。通过LSD和O-H-LSD两种方法对先前通过气相色谱-质谱法发现含有LSD的68个人类尿液样本进行了重新分析。这些样品的O-H-LSD平均浓度约为LSD浓度的16倍。因为两种液相色谱方法都能产生相似的结果,所以两种方法都可以很容易地适用于O-H-LSD分析,以用于大量的药物测试实验室。此外,通过针对主要的LSD代谢物进行分析来显着增加LSD检测时间窗口的可能性,可能会影响其他无毒工作场所程序来测试LSD。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号