首页> 外文期刊>Journal of Aerosol Science >Inter-comparison of optical absorption coefficients of atmospheric aerosols determined by a multi-wavelength photoacoustic spectrometer and an Aethalometer under sub-urban wintry conditions
【24h】

Inter-comparison of optical absorption coefficients of atmospheric aerosols determined by a multi-wavelength photoacoustic spectrometer and an Aethalometer under sub-urban wintry conditions

机译:多波长光声光谱仪和热度计在郊区寒冬条件下确定的大气气溶胶光吸收系数的比对

获取原文
获取原文并翻译 | 示例
           

摘要

The wavelength-dependent optical absorption coefficients (OACs) of atmospheric aerosols were measured at a sub-urban site in Budapest under wintry conditions simultaneously with a commercially available, seven-wavelength (7λ) Aethalometer instrument, and with a recently developed four-wavelength (1064,532,355 and 266 nm) photoacoustic spectrometer (4λ-PAS). While there is a good agreement between the readings on the two systems in the near-infrared and in the visible, the OACs measured by the 4λ-PAS in the UV region were systematically higher than the OACs extrapolated to the UV region from data measured in the visible by the Aethalometer, and the relative deviation between the readings of the instruments in the UV was found to increase exponentially as a function of the Angstrom exponent fitted to the wavelength dependency measured by either system. This result indicates that the 4λ-PAS has more diagnostic value than the Aethalometer, i.e. it has higher sensitivity to discriminate between different aerosol sources. Furthermore, our finding can significantly affect the results of radiative transfer calculations for atmospheric aerosols, in which wavelength-dependent optical absorption, usually measured by Aethalometers, is an important input parameter.
机译:在寒冷的条件下,在布达佩斯郊区的一个站点上,使用市售的七波长(7λ)测速仪和最近开发的四波长(AAC)仪,测量了大气气溶胶的波长依赖性光吸收系数(OAC)。 1064,532,355和266 nm)光声光谱仪(4λ-PAS)。虽然在近红外和可见光两个系统的读数之间有很好的一致性,但是4λ-PAS在紫外区域测得的OAC系统地高于从红外测得的数据外推到紫外区域的OAC。可以通过湿度计观察到可见光,并且发现紫外线下仪器读数之间的相对偏差呈指数增加,这是根据任一系统测得的波长依赖性拟合的Angstrom指数的函数。该结果表明,4λ-PAS比起湿度计具有更高的诊断价值,即,它具有更高的区分不同气溶胶来源的灵敏度。此外,我们的发现会显着影响大气气溶胶的辐射传递计算结果,在该计算中,通常由Aethalometers测量的与波长相关的光吸收是重要的输入参数。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号