The Couette centrifugal particle mass analyzer(CPMA)classifies particles by their mass-to-charge ratio.Unlike the aerosol particle mass(APM)analyzer,the Couette CPMA uses a stable system of forces to improve the transfer function of the classifier.A prototype Couette CPMA has been built and tested.The experimental results from the prototype agree well with theory and it is found that indeed the transfer function of the Couette CPMA is better than the APM's.Experimental work is shown using a differential mobility analyzer(DMA)and Couette CPMA to classify polystyrene latex(PSL)and di-2-ethylhexyl seba-cate(DEHS)particles.By measuring the mass of PSL particles the absolute uncertainty of the Couette CPMA was found to be 6.7-38% higher in terms of mass(or 2.2-11 % higher in terms of equivalent diameter)than the expected value.However,when the DMA-CPMA system was calibrated with PSL particles the density of DEHS particles was measured to within approximately 3% of the expected value.
机译:Couette离心颗粒质量分析仪(CPMA)通过质荷比对颗粒进行分类。与气溶胶颗粒质量(APM)分析仪不同,Couette CPMA使用稳定的力系统来改善分类器的传递功能。 Couette CPMA已构建并经过测试,原型的实验结果与理论吻合良好,发现Couette CPMA的传递函数确实优于APM的传递函数。使用差动迁移率分析仪(DMA)和Couette CPMA对聚苯乙烯乳胶(PSL)和二-2-乙基己基癸二酸酯(DEHS)颗粒进行分类。通过测量PSL颗粒的质量,Couette CPMA的绝对不确定度比质量高6.7-38% (或等效直径的2.2-11%)高于预期值。但是,当用PSL颗粒校准DMA-CPMA系统时,测得的DEHS颗粒密度在预期值的3%之内。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号