We present a study of the laboratory version of a time-resolved photoacoustic mammoscope built based on a parallel plate geometry. An intralipid (IL) phantom with embedded cylindrical dyed gel inhomogeneities of different sizes and optical absorptions was used to test the quality of 3D image reconstruction and ability of studied instrument to detect optical contrast between imaged objects. The temporal backward projection Fourier transform based method was used for image reconstruction. The results of this study show that width at half maximum of images obtained by the photoacoustic mammoscope correlate well with lateral profile of imaged structures and that the optical contrast between imaged objects can be measured with deviation of 35%.
机译:我们目前对基于平行板几何结构的时间分辨光声乳房镜的实验室版本进行研究。使用具有不同大小和光学吸收度的嵌入式圆柱形染色凝胶不均匀性的脂内(IL)体模来测试3D图像重建的质量以及所研究仪器检测成像对象之间光学对比度的能力。基于时间向后投影傅立叶变换的方法用于图像重建。这项研究的结果表明,通过光声乳房镜获得的图像的最大一半宽度与成像结构的侧面轮廓密切相关,并且可以测量成像对象之间的光学对比度,偏差为35%。