BACKGROUND: Atrioventricular nodal reentrant tachycardia (AVNRT) is a frequent supraventricular tachycardia in children and young adults. Despite favourable success rates of catheter ablation, conventional fluoroscopic catheter guidance is associated with risks of low-dose ionizing radiation for the patient and the personnel. Here we describe a technique for zero-fluoroscopy catheter ablation using contact force technology. PATIENTS AND METHODS: Zero-fluoroscopy catheter ablation was attempted in 12 patients with AVNRT (median age 20 years; range 11-75 years). An ablation catheter with integrated contact force sensor and a nonfluoroscopic electroanatomical mapping system was used for visualization of cardiovascular structures. Mean contact forces during mapping and ablation were restricted to an upper limit of 50 g to avoid cardiovascular injuries. RESULTS: Zero-fluoroscopy catheter ablation was performed successfully and uneventfully in all patients. There were no arrhythmia recurrences during a median follow-up of 6.2 months (range 2.7-12.8). CONCLUSION: Zero-fluoroscopy catheter ablation of AVNRT is possible and appears simple yet safe, when a nonfluoroscopic electroanatomical mapping system is used in combination with an ablation catheter with integrated contact force sensor. The presented technique could thus be easily employed in most electrophysiological laboratories.
机译:背景:房室结折返性心动过速(AVNRT)是儿童和年轻人中常见的室上性心动过速。尽管导管消融的成功率很高,但是常规的荧光镜导管引导仍会给患者和人员带来低剂量电离辐射的风险。在这里,我们描述了一种使用接触力技术的零荧光镜导管消融技术。患者与方法:12例AVNRT患者(中位年龄20岁;范围11-75岁)尝试进行零荧光透视导管消融术。具有集成接触力传感器和非荧光镜电解剖标测系统的消融导管用于可视化心血管结构。制图和消融期间的平均接触力被限制在50 g的上限,以避免心血管损伤。结果:所有患者均成功进行了零荧光透视导管消融治疗。中位随访6.2个月(2.7-12.8),无心律失常复发。结论:当将非荧光镜电解剖标测系统与带有集成接触力传感器的消融导管结合使用时,AVNRT的零荧光镜消融术是可行的,并且看上去简单而安全。因此,所提出的技术可以容易地在大多数电生理实验室中采用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号