A comparative evaluation of cone beam CT and micro-CT on trabecular bone structures in the human mandible.

机译:锥形束CT和微型CT对人类下颌骨小梁骨结构的比较评估。

获取原文
获取原文并翻译 | 示例
           

摘要

The main purpose of this study was to determine the accuracy of cone beam CT (CBCT) in measuring the trabecular bone microstructure, in comparison with micro-CT. The subobjective was to examine to what extent bone quality assessment is influenced by X-ray tube current and voltage settings as well as soft tissue surrounding the bone.Eight human mandibular bone samples were scanned using three different clinical exposure protocol within water (W1-3) and without water (NW1-3) by a high-resolution (80 μm) CBCT machine (3D Accuitomo 170(?); Morita, Kyoto, Japan). Subsequently, the samples underwent micro-CT scanning (SkyScan 1174?; SkyScan, Antwerp, Belgium). After image acquisition, similar volumes of interest of the trabecular structures captured with CBCT and micro-CT were aligned with each other. Segmentation was then performed, and the morphometric parameters were quantified within the volumes of interest by CTAn software (CTAnalyser(?); SkyScan, Antwerp, Belgium). Descriptive statistical analyses and multiple comparisons between all protocols were applied in R software.High positive Pearson's correlation coefficients were observed between CBCT and micro-CT protocols for all tested morphometric indices except for trabecular thickness. No significant differences were observed between all exposure protocols except for trabecular separation. When examining the soft-tissue effect on trabecular bone structures, no significant differences between NW (1-3) and W (1-3) protocols were observed for all variables.The present study demonstrated the potential of high-resolution CBCT imaging for in vivo applications of quantitative bone morphometry and bone quality assessment. However, the overestimation of morphometric parameters and acquisition settings in CBCT must be taken into account.
机译:这项研究的主要目的是确定与微型CT相比,锥形束CT(CBCT)在测量小梁骨微结构中的准确性。目的是检查X射线管电流和电压设置以及骨骼周围的软组织在多大程度上影响骨质量评估。使用三种不同的临床暴露方案在水中扫描八个人下颌骨样品(W1-3 ),并且使用高分辨率(80μm)CBCT机器(3D Accuitomo 170(?);日本京都森田市)不加水(NW1-3)。随后,对样品进行微型CT扫描(SkyScan 1174 ?;比利时安特卫普的SkyScan)。图像采集后,用CBCT和micro-CT捕获的相似大小的骨小梁结构相互对齐。然后进行分割,并通过CTAn软件(CTAnalyser(?); SkyScan,安特卫普,比利时)在感兴趣的体积内对形态计量学参数进行定量。在R软件中进行了描述性统计分析和所有方案之间的多重比较。除小梁厚度外,对于所有测试的形态计量指标,CBCT和micro-CT方案之间均观察到高正Pearson相关系数。除小梁分离外,所有暴露方案之间均未观察到显着差异。当检查软组织对小梁骨结构的影响时,在所有变量中,NW(1-3)和W(1-3)方案之间均未观察到显着差异。本研究证明了高分辨率CBCT成像在小梁骨中的潜力。体内应用定量骨形态测量和骨质量评估。但是,必须考虑CBCT中形态参数和采集设置的过高估计。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号