首页> 外文期刊>DentoMaxilloFacial Radiology >Radiofrequency heating and magnetically induced displacement of dental magnetic attachments during 3.0 T MRI
【24h】

Radiofrequency heating and magnetically induced displacement of dental magnetic attachments during 3.0 T MRI

机译:3.0 T MRI期间射频加热和磁致牙齿附件的位移

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

The aim of this study was to estimate the risk of injury from dental magnetic attachments due to their radiofrequency (RF) heating and magnetically induced displacement during 3.0 T MRI. Methods: To examine the magnetic attachments, we adopted the American Society for Testing and Materials F2182-02a and F2052-06 standards in two MRI systems (Achieva 3.0 T Nova Dual; Philips, Tokyo, Japan, and Signa HDxt 3.0 T; GE Healthcare, Milwaukee, WI). The temperature change was measured in a cylindrical keeper (GIGAUSS D600; GC, Tokyo, Japan) with coping of the casting alloy and a keeper with a dental implant at the maximum specific absorption rate (SAR) for 20 min. To measure the magnetically induced displacement force, three sizes of keepers (GIGAUSS D400, D600 and D1000) were used in deflection angle tests conducted at the point of the maximum magnetic field strength. Results: Temperature elevations of both coping and implant were higher in the Signa system than in the Achieva system. The highest temperature changes in the keeper with implant and keeper with coping were 0.6 °C and 0.8 °C in the Signa system, respectively. The temperature increase did not exceed 1.0 °C at any location. The deflection angle (a) was not measurable because it exceeded 90°. GIGAUSS D400 required an extra 3.0 g load to constrain the deflection angle to less than 45° GIGAUSS D600 and D1000 required 5.0 and 9.0 g loads, respectively.
机译:这项研究的目的是估计在3.0 T MRI期间由于其磁性(RF)加热和磁感应位移而引起的牙科磁性附件受伤的风险。方法:为了检查磁性附件,我们在两个MRI系统(Achieva 3.0 T Nova Dual,日本东京的Philips,Signa HDxt 3.0 T,GE Healthcare的两个MRI系统)中采用了美国测试和材料学会F2182-02a和F2052-06标准。 ,威斯康星州密尔沃基)。在圆柱形保持器(GIGAUSS D600; GC,东京,日本)中以铸件合金的顶盖和保持器与牙科植入物以最大比吸收率(SAR)测量20分钟的温度变化。为了测量磁感应位移力,在最大磁场强度点进行的偏转角测试中使用了三种尺寸的保持架(GIGAUSS D400,D600和D1000)。结果:Signa系统中的应对和植入物的温度升高均高于Achieva系统。在Signa系统中,带植入物的支架和带顶帽的支架的最高温度变化分别为0.6°C和0.8°C。任何位置的温度升高均不超过1.0°C。偏转角(a)超过90°,因此无法测量。 GIGAUSS D400需要额外的3.0 g载荷,以将偏转角限制为小于45°。GIGAUSS D600和D1000分别需要5.0 g和9.0 g载荷。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号