Chemical exchange saturation transfer MR imaging of articular cartilage glycosaminoglycans at 3 T: Accuracy of B_0 Field Inhomogeneity corrections with gradient echo method

机译:3 T下关节软骨糖胺聚糖的化学交换饱和转移MR成像:梯度回波法校正B_0场的不均匀性

获取原文
获取原文并翻译 | 示例

摘要

Glycosaminoglycan Chemical Exchange Saturation Transfer (gagCEST) is an important molecular MRI methodology developed to assess changes in cartilage GAG concentrations. The correction for B_0 field inhomogeneity is technically crucial in gagCEST imaging. This study evaluates the accuracy of the B_0 estimation determined by the dual gradient echo method and the effect on gagCEST measurements. The results were compared with those from the commonly used z-spectrum method. Eleven knee patients and three healthy volunteers were scanned. Dual gradient echo B_0 maps with different ΔTE values (1, 2, 4, 8, and 10 ms) were acquired. The asymmetry of the magnetization transfer ratio at 1 ppm offset referred to the bulk water frequency, MTRasym(1 ppm), was used to quantify cartilage GAG levels. The B0 shifts for all knee patients using the z-spectrum and dual gradient echo methods are strongly correlated for all ΔTE values used (r = 0.997 to 0.786, corresponding to ΔTE = 10 to 1 ms). The corrected MTRasym(1 ppm) values using the z-spectrum method (1.34% ± 0.74%) highly agree only with those using the dual gradient echo methods with ΔTE = 10 ms (1.72% ± 0.80%; r = 0.924) and 8 ms (1.50% ± 0.82%; r = 0.712). The dual gradient echo method with longer ΔTE values (more than 8 ms) has an excellent correlation with the z-spectrum method for gagCEST imaging at 3 T.
机译:糖胺聚糖化学交换饱和转移(gagCEST)是一种重要的分子MRI方法,旨在评估软骨GAG浓度的变化。 B_0场不均匀性的校正在gagCEST成像中在技术上至关重要。这项研究评估了由双重梯度回波方法确定的B_0估计的准确性以及对gagCEST测量的影响。将结果与常用z光谱方法的结果进行比较。扫描了11名膝盖患者和3名健康志愿者。获取具有不同ΔTE值(1、2、4、8和10 ms)的双梯度回波B_0映射。相对于总水频率MTRasym(1 ppm),在偏移1 ppm时磁化传递比率的不对称性用于量化软骨GAG含量。使用z频谱和双重梯度回波方法的所有膝关节患者的B0位移与所用的所有ΔTE值都具有高度相关性(r = 0.997至0.786,对应于ΔTE= 10至1 ms)。使用z光谱方法校正的MTRasym(1 ppm)值(1.34%±0.74%)仅与使用双梯度回波方法的ΔTE= 10 ms(1.72%±0.80%; r = 0.924)和8高度一致ms(1.50%±0.82%; r = 0.712)。 ΔTE值较长(大于8毫秒)的双梯度回波方法与z光谱方法在3 T时进行gagCEST成像具有极好的相关性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号