摘要

Objective Characterization of magnetic susceptibility artefacts with assessment of the gradient-echo signal decay function of echo time, pixel size, and object geometry in the case of air-filled cylinders embedded in water.Materials and methods Experiments were performed with a 0.2 T magnet on a network of small interacting air-filled cylinders along with Magnetic resonance imaging (MRI) simulations integrating intravoxel dephasing. Signal decay over echo time was assessed at different pixel sizes on real and simulated images. The effects of radius, distance between cylinders and main magnetic field were studied using simulation.
机译:目的对磁化伪像进行表征,并评估在空气中装有空气的圆柱体情况下回波时间,像素大小和物体几何形状的梯度回波信号衰减函数。材料和方法在0.2 T磁体上进行了实验一个互动的小充气气瓶网络以及整合了体素内移相的磁共振成像(MRI)模拟。在真实和模拟图像上,在不同像素大小下评估了回波时间内的信号衰减。通过仿真研究了半径,圆柱体之间的距离和主磁场的影响。