Measurements of left ventricular dimensions using real-time acquisition in cardiac magnetic resonance imaging: comparison with conventional gradient echo imaging

机译:在心脏磁共振成像中使用实时采集测量左心室尺寸:与常规梯度回波成像的比较

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

This study investigates the use of real-time acquisition in cardiac magnetic resonance imaging (MRI) for measurements of left ventricular dimensions in comparison with conventional gradient echo acquisition. Thirty-one subjects with a variety of left ventricular morphologies to represent a typical clinical population were studied. Short-axis data sets of the left ventricle (LV) were acquired using a conventional turbo-gradient echo and an ultrafast hybrid gradient echo/echo planar sequence with acquisition in real-time. End-diastolic volume (EDV), end-systolic volume (ESV), ejection fraction (EF) and left ventricular mass (LV mass) were measured. The agreement between the two acquisitions and interobserver, intraobserver and interstudy variabilities were determined. The bias between the two methods was 5.86 ml for EDV, 0.23 ml for ESV and 0.94% for EF. LV mass measurements were significantly lower with real-time method (mean bias 14.38 g). This is likely to be the result of lower spatial resolution and chemical shift artefacts with the real-time method. Interobserver, intraobserver and interstudy variabilities were low for all parameters. In conclusion, real time acquisition in MRI can provide accurate and reproducible measurements of LV dimensions in subjects with normal as well as abnormal LV morphologies, but LV mass measurements were lower than with conventional gradient echo imaging.
机译:本研究调查了与常规梯度回波采集相比,实时采集在心脏磁共振成像(MRI)中用于测量左心室尺寸的用途。研究了代表各种典型临床人群的具有各种左心室形态的31名受试者。左心室(LV)的短轴数据集使用常规涡轮梯度回波和超快混合梯度回波/回波平面序列进行实时采集。测量舒张末期容积(EDV),收缩末期容积(ESV),射血分数(EF)和左心室质量(LV质量)。确定了两次采集之间的一致性,以及观察者之间,观察者内部和研究之间的差异。两种方法之间的偏差为EDV为5.86 ml,ESV为0.23 ml,EF为0.94%。实时法测得的左心室质量明显降低(平均偏倚14.38 g)。这可能是由于实时方法降低了空间分辨率和化学位移伪像的结果。所有参数的观察者间,观察者内和研究间变异性均较低。总之,MRI的实时采集可以为正常和异常LV形态的受试者提供准确且可重复的LV尺寸测量结果,但LV质量测量值低于常规梯度回波成像。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号