This paper presents an automatic method of correcting non-uniform RF coil response for the classification of body composition using MR imaging. By linear mosaic modelling, the smoothly but non-linearly varying bias field, which modulates tissue intensities within the image, was corrected. The overlapping between adjacent mosaics ensured consistent segmentation of body fat content and the effectiveness of the technique was validated by both phantom and in vivo experiments. Ten whole body composition data sets, each with 39 trans-axial slices, were acquired. Automatic segmentation results using the proposed technique were compared with those from manual delineations. The automatic segmentation method was found to be highly accurate and the mean percentage error between the two methods was less than 1.5%.
机译:本文提出了一种自动校正非均匀射频线圈响应的方法,用于使用MR成像对人体成分进行分类。通过线性镶嵌建模,可以校正平滑但非线性变化的偏置场,该偏置场可调节图像内的组织强度。相邻镶嵌图之间的重叠确保了人体脂肪含量的一致分割,并且该技术的有效性已通过幻像和体内实验验证。获得十个全身成分数据集,每个数据集包含39个跨轴切片。使用提出的技术的自动分割结果与手动描绘的结果进行了比较。发现自动分割方法非常准确,并且两种方法之间的平均百分比误差小于1.5%。