Automated oscillometric determination of the ankle-brachial index provides accuracy necessary for office practice.

机译:脚腕臂指数的自动示波法测定提供了办公实践所需的准确性。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Peripheral arterial disease (PAD) remains underdiagnosed by primary care and cardiovascular physicians. The office-based assessment of PAD is limited by the need for specialized equipment and the time required for performance of the ankle-brachial index (ABI). We explored whether the accuracy of automated ABI measurement by oscillometry compared favorably with the gold-standard method using continuous-wave Doppler ultrasound. Consecutive patients referred to our university hospital noninvasive vascular laboratory for ABI measurement were invited for participation. Of 205 patients, 201 participated, including 55 with PAD. The ABI was measured by automated oscillometry and Doppler ultrasound. The test of trends revealed a correlation coefficient of 0.78 in the left leg and 0.78 in the right leg (P<0.01 for both). The mean ABI difference between methods was 0.04+/-0.01 and 0.06+/-0.01, respectively, in the left and right legs. The differences between the methods followed a normal distribution. Oscillometric determination of the ABI provides an accurate determination of the ABI in an outpatient population. Our findings show automated oscillometry to be a reliable and easier method of ABI measurement, lowering the barrier to incorporation of this diagnostic test into clinical practice.
机译:初级保健和心血管医师仍未充分诊断周围动脉疾病(PAD)。基于办公室的PAD评估受限于对专用设备的需求以及执行踝肱指数(ABI)所需的时间。我们探讨了使用示波法自动测量ABI的准确性是否与使用连续波多普勒超声的金标准方法相比是否令人满意。邀请转诊至我院大学无创血管实验室进行ABI测量的患者参加。在205位患者中,有201位患者参加,包括55位PAD患者。通过自动示波法和多普勒超声测量ABI。趋势检验显示左腿的相关系数为0.78,右腿的相关系数为0.78(两者均P <0.01)。左腿和右腿方法之间的平均ABI差异分别为0.04 +/- 0.01和0.06 +/- 0.01。方法之间的差异遵循正态分布。示波法测定ABI可准确确定门诊患者的ABI。我们的发现表明,自动示波法是一种可靠且更容易的ABI测量方法,从而降低了将该诊断测试纳入临床实践的障碍。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号