首页> 外文期刊>Hip international: the journal of clinical and experimental research on hip pathology and therapy >A novel device to measure acetabular inclination with patients in lateral decubitus
【24h】

A novel device to measure acetabular inclination with patients in lateral decubitus

机译:一种测量外侧卧位患者髋臼倾角的新型装置

获取原文
获取原文并翻译 | 示例
       

摘要

The ability to measure acetabular cup orientation accurately during total hip arthroplasty represents a significant challenge. The aim of this research was to develop and evaluate a novel low cost mechanical device for measuring operative acetabular inclination. Cup implantation was simulated in two trials using the novel device: firstly involving surgeons and engineers orientating acetabular cups with sawbone pelves at a range of inclination angles (20°-55° in 5° increments); secondly in a simulated intra-operative scenario with surgeons. Target angles were compared with achieved angles and deviations from desired angles were recorded. In addition, all participants orientated cups under the same conditions using two other techniques: freehand and with a propriatory Mechanical Alignment Guide. In the first trial, the mean errors (deviations) using freehand technique, the mechanical alignment guide and the new device were 5.2° +/- 4.3° (range 0.1-22.0), 3.6° +/- 3.9° (range 0.1°-33.6°) and0.5° +/- 0.4° (range 0.0-1.9) respectively. In the second trial, the mean error for freehand technique,mechanical alignment guide and the new device were 6.2° +/- 4.2° (range 0.2-18.2), 3.8° +/- 3.3°(range 0.0-19.1) and 0.6° +/- 0.5° (range 0.0-1.8) respectively. The new device has the potential to allowthe surgeon to choose and record operative inclination accurately during total hip arthroplasty in the lateral decubitus position.
机译:在全髋关节置换术中准确测量髋臼杯方向的能力代表了一项重大挑战。这项研究的目的是开发和评估一种新型的低成本机械装置,用于测量手术髋臼倾斜度。使用该新型装置在两项试验中模拟了杯的植入:首先,由外科医生和工程师参与,在一定的倾斜角度范围内(20°-55°,以5°增量)将髋臼杯与锯齿骨定向。其次是在有外科医师的模拟术中情况下。将目标角度与获得的角度进行比较,并记录与所需角度的偏差。此外,所有参与者都使用其他两种技术在相同条件下定向杯子:徒手绘制和带有专有的机械对准指南。在第一次试验中,使用徒手技术,机械对准指南和新设备的平均误差(偏差)为5.2°+/- 4.3°(范围0.1-22.0),3.6°+/- 3.9°(范围0.1°- 33.6°)和0.5°+/- 0.4°(范围0.0-1.9)。在第二次试验中,徒手画法,机械对准指南和新设备的平均误差为6.2°+/- 4.2°(范围0.2-18.2),3.8°+/- 3.3°(范围0.0-19.1)和0.6°分别为+/- 0.5°(范围为0.0-1.8)。新设备有可能允许外科医生在侧卧位全髋关节置换术期间准确选择并记录手术倾斜度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号