首页> 外文期刊>Haemostasis >Comparative study of a portable monitor for prothrombin time determination, Coaguchek, with three systems for control of oral anticoagulant treatment.
【24h】

Comparative study of a portable monitor for prothrombin time determination, Coaguchek, with three systems for control of oral anticoagulant treatment.

机译:便携式凝血酶原时间测定仪Coaguchek与三种控制口服抗凝治疗的系统的比较研究。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Self-testing of oral anticoagulation is a new possibility related to the development of portable capillary whole blood prothrombin time monitors. The aim of this study was to evaluate one of this monitors, Coaguchek, with respect to its comparability with our routine prothrombin time determination system, as well as with the reference manual technique and two thromboplastins of high sensitivity, Manchester Reagent and one manufactured in our center, Thromboplastin Bilbao, in a group of patients on oral anticoagulant treatment. Although a correlation of r = 0.9271 was found between international normalized ratio (INR) values of Coaguchek and our routine method, Neoplastine/STA analyzer, the difference of the INR scatter increased with the magnitude of measurement, being lowest for INR between the portable monitor and Manchester Reagent and Thromboplastin Bilbao, with a similar coefficient of correlation, r = 0.8948 and r = 0.8905, respectively. A test was performed showing a 65.6% agreement with the INR values of the STA analyzer, 66.4% with Manchester Reagent and 73.4% with Thromboplastin Bilbao. On the basis of this correspondence with laboratory prothrombin time results Coaguchek may be considered as a possible option for monitoring anticoagulated patients even though patients should be given instructions and advice as regards the management and interpretation of the results.
机译:口服抗凝剂的自我测试是与便携式毛细管全血凝血酶原时间监测仪的开发有关的新可能性。这项研究的目的是评估其中一台监测仪Coaguchek与我们常规的凝血酶原时间测定系统,参考手册技术和两种高灵敏度凝血酶原蛋白(曼彻斯特试剂)的可比性。中心,毕尔巴鄂凝血活酶,在一组接受口服抗凝治疗的患者中。尽管在Coaguchek的国际标准化比率(INR)值和我们的常规方法Neoplastine / STA分析仪之间发现r = 0.9271的相关性,但INR散射的差异随测量幅度的增加而增加,对于便携式显示器之间的INR最低相关系数分别为r = 0.8948和r = 0.8905。进行的测试显示,与STA分析仪的INR值具有65.6%的一致性,与Manchester试剂具有66.4%的一致性,与毕尔巴鄂促凝血酶原蛋白的73.4%的一致性。基于这种与实验室凝血酶原时间结果的对应关系,即使应该就结果的管理和解释给予患者指导和建议,也可以将Coaguchek视为监测抗凝患者的可能选择。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号