This study tested the possible damage to the internal implant connection provoked by repeated disconnection and reconnection of prosthetic components. Using a light-structured scanner, connection deformation was inferred by three-dimensional (3D) positional changes of a "reference" abutment before and after multiple dis- and reconnections. Measurements were taken after 1, 2, 5, 10, and 20 insertions of titanium abutments in 16 internal hexagon implants. Statistical analysis revealed that multiple dis- and reconnections could cause a deformation in the implant-abutment connection, proportional to the number of insertion procedures. However, below the threshold of 10 cycles, 3D deformation values were minimal.
机译:这项研究测试了假体组件的反复断开和重新连接可能引起的内部植入物连接的损坏。使用光结构扫描仪,通过多次断开和重新连接之前和之后“参考”基台的三维(3D)位置变化,可以推断出连接变形。在16个内部六角形种植体中插入1、2、5、10和20钛基台后进行测量。统计分析表明,多次断开和重新连接可能会导致植入物-基台连接变形,与插入程序的数量成正比。但是,在10个循环的阈值以下,3D变形值最小。