Effects of a cementing technique in addition to luting agent on the uniaxial retention force of a single-tooth implant-supported restoration: an in vitro study.

机译:除浸润剂外,固井技术对单齿种植体支持的修复体单轴固位力的影响:一项体外研究。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Excess residual cement around the implant margin has been shown to be detrimental to the peri-implant tissue. This in vitro study examines the retentive strengths of two different cementing techniques and two different luting agents on a machined titanium abutment and solid screw implants. The amount of reduction of excess cement weight between the two cementation techniques was assessed. Forty gold castings were fabricated for 4.1 mm in diameter and 10 mm in length solid-screw dental implants paired with 5.5-mm machined titanium abutments. Twenty implants received a provisional cement, and 20 implants received a definitive cement. Each group was further divided into two groups. In the control group, cement was applied and the castings seated over the implant-abutment assembly. The excess cement was then removed. In the study group, a "practice abutment" was used to express excess cement prior to cementation. The weight of the implant-casting assembly was measured and the residual weight of cement was calculated. The samples were then stored for 24 hours at 100% humidity prior to tensile strength testing. Statistical analysis revealed significant differences in tensile strength across the groups. Further Tukey tests showed no significant difference in tensile strength between the practice abutment technique and the conventional technique for both definitive and provisional cements. There was a significant reduction in residual cement weight, irrespective of the type of cement, when the practice abutment was used prior to cementation. Cementation of implant restorations on a machined abutment using the practice abutment technique and definitive cement may provide similar uniaxial retention force and significantly reduced residual cement weight compared to the conventional technique of cement removal.
机译:种植体边缘周​​围多余的残留骨水泥已被证明对种植体周围组织有害。这项体外研究检查了在机械加工的钛基台和实心螺钉植入物上两种不同的固井技术和两种不同的浸润剂的保持力。评估了两种胶结技术之间多余水泥重量的减少量。制作了四十个金铸件,直径为4.1毫米,长度为10毫米,与5.5毫米机加工的钛基台配对。 20个植入物接受了临时胶合剂,而20个植入物接受了确定的胶合剂。每一组进一步分为两组。在对照组中,应用水泥,并将铸件放置在种植体-基台组件上。然后除去多余的水泥。在研究组中,在粘结之前,使用“练习基台”来表示多余的水泥。测量植入物铸造组件的重量,并计算出水泥的残余重量。然后在拉伸强度测试之前,将样品在100%湿度下存储24小时。统计分析表明,各组的拉伸强度存在显着差异。进一步的Tukey测试表明,确定性和临时性水泥的实际桥台技术和常规技术之间的拉伸强度没有显着差异。当在胶凝之前使用练习基台时,无论胶凝剂的类型如何,残余的水泥重量都会显着降低。与常规的牙结石去除技术相比,使用实践基牙技术和确定性牙结石将种植体修复物固结在机械基牙上可以提供相似的单轴固位力,并显着降低了残余牙结石的重量。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号