It has been hypothesized that a correlation exists between the density of surrounding cortical bone and the stability of an implant under percussion loading that can be used to quantify the implant's osseointegration. The purpose of the present research was to explore whether quantitative percussion testing of dental implants gives reasonable indications of the level of osseointegration that are consistent with bone configuration and its influence on osseointegration quality.Data from percussion testing of a live human subject, obtained using the Periometer, were compared with corresponding bone density estimates from high-resolution computed tomography images and postmortem percussion probe data.The results confirm the hypothesis that the nature of an implant's response to percussion is determined by its cortical bone support.The findings suggest that the cortical bone supporting the crestal and apical regions of the implant is primarily responsible for structural stability.
机译:假设在冲击载荷下,周围皮质骨的密度与植入物的稳定性之间存在相关性,该相关性可用于量化植入物的骨整合。本研究的目的是探讨牙种植体的定量打击测试是否能给出合理的骨整合水平指标,该指标与骨骼结构及其对骨整合质量的影响是一致的。将视野仪与高分辨率计算机断层扫描图像和死后敲击探针数据得到的相应骨密度估计值进行比较,结果证实了假说,即植入物对敲击反应的性质取决于其皮层骨支持物。支撑植入物的顶部和顶部区域的骨骼主要负责结构稳定性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号