Digitizing implant position locators on master casts: comparison of a noncontact scanner and a contact-probe scanner.

机译:数字化铸件上的植入物位置定位器:非接触式扫描仪和接触式探针扫描仪的比较。

获取原文
获取原文并翻译 | 示例
       

摘要

Although several investigations of the fit of computer-aided design/computer-assisted manufactured implant restorations have been reported in the literature, very limited information is available on the accuracy of dental scanners for surface data acquisition. The objective of this research project was to investigate the accuracy and repeatability of a three-dimensional noncontact laser scanning system based on conoscopic holography recording the spatial orientation of an implant in vitro.One master cast containing five nonparallel implant replicas with scan abutments (implant position locator [IPL], Nobel Biocare) was repeatedly digitized by two commercially available dental scanners and an industrial highprecision coordinate measuring machine as a reference. Repeated measurements of the center point and vector analysis were applied to determine the accuracy and repeatability of IPL orientation. The Wilcoxon signed rank test modified for equivalence was applied to determine whether the systems were equally accurate based on given threshold limits of 0.010, 0.005, and 0.001 mm.All between-system tests identified significant differences, although most differences were very small in absolute terms. Tests for equivalence based on difference windows were exploratorily significant for 0.010 and 0.005 mm.Conoscopic holography scan technology is a valid option to supplement dental touch-probe scanners for laboratory digitization of implant master casts.
机译:尽管在文献中已经报道了对计算机辅助设计/计算机辅助制造的植入物修复体的适合性的一些研究,但是关于用于表面数据采集的牙科扫描仪的准确性的信息非常有限。本研究项目的目的是研究基于锥光全息术的三维非接触式激光扫描系统的准确性和可重复性,该系统记录体外植入物的空间取向。一个主铸件包含五个带有扫描基台的非平行植入物复制品(植入物位置定位器[IPL],Nobel Biocare)被两台市售的牙科扫描仪和一台工业高精度坐标测量机作为参考重复进行了数字化处理。重复测量中心点并进行矢量分析,以确定IPL方向的准确性和可重复性。根据给定的阈值限制0.010、0.005和0.001 mm,使用经过修改以等效的Wilcoxon符号秩检验来确定系统是否同样准确。所有系统间测试均确定了显着差异,尽管大多数差异在绝对意义上很小。基于差异窗口的等效性测试对于0.010和0.005 mm来说具有显着的意义。锥形全息扫描技术是补充牙科触摸探针扫描仪以对植入物母模进行实验室数字化的有效选择。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号