In vivo lactate measurement in human tissue by near-infrared diffuse reflectance spectroscopy

机译:近红外漫反射光谱法在人体组织中体内乳酸的测量

获取原文
获取原文并翻译 | 示例
           

摘要

We have evaluated the potential of near-infrared spectroscopy(NIRS)as a technique for in vivo measurement of fatigue during exercise in humans.By using 2D correlation spectroscopy,lactate was identified as the primary specie that was monitored by the in vivo measurements.Estimates and calibration of the lactate concentration by NIRS was made using partial least squares(PLS)regression analysis and leave-N-out cross-validation on second derivative spectra.Regression analysis provided a correlation coefficient of 0.74 and a standard error of cross-validation(SECV)of 2.21 mmol/1.To study the parameters that impact the spectra baseline and the correlation between the calculated model and the data,lactate spectra were referenced against a baseline spectrum for each individual.For the referenced data,a correlation coefficient of 0.97 and a standard error of cross-validation of 0.76 mmol/1 between the enzymatic assay and the NIRS technique were found.The results suggest that NIRS may provide a valuable tool to assess in vivo physiological status for both research and clinical needs.
机译:我们已经评估了近红外光谱(NIRS)作为体内测量人体运动中疲劳的技术的潜力。通过使用2D相关光谱法,乳酸盐被鉴定为通过体内测量监测的主要物种。使用偏最小二乘回归分析和二阶导数光谱的离开-N-交叉验证对NIRS进行乳酸浓度校准。回归分析提供的相关系数为0.74,交叉验证的标准误为( SECV)为2.21 mmol / 1。为研究影响光谱基线的参数以及计算模型与数据之间的相关性,将乳酸谱与每个个体的基线光谱进行对比。对于参考数据,相关系数为0.97酶促测定与NIRS技术之间的交叉验证标准误为0.76 mmol / 1。评估研究和临床需求的体内生理状态的宝贵工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号