首页> 外文期刊>Veterinary Ophthalmology >Comparison of intraocular pressure measurements between the Tono-Pen XL (R) and Perkins (R) applanation tonometers in dogs and cats
【24h】

Comparison of intraocular pressure measurements between the Tono-Pen XL (R) and Perkins (R) applanation tonometers in dogs and cats

机译:狗和猫的Tono-Pen XL(R)和Perkins(R)压平眼压计之间眼压测量值的比较

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Objective To compare the accuracy between two applanation tonometers, Tono-Pen XL (R) and Perkins (R), in ophthalmoscopically normal dogs and cats. Animals Both eyes of 25 conscious and healthy dogs and cats were evaluated. Both eyes of five dogs and cats immediately after sacrifice were used as controls for the postmortem study. Procedure In conscious animals, the tonometry was performed with topical anesthesia using 0.5% proxymetacaine eye drops for both tonometers and 1% fluorescein eye drops for the Perkins tonometer. Readings of intraocular pressure (IOP) in the postmortem study were taken using manometry and tonometry by Tono-Pen XL (R) and Perkins (R). Results The correlation coefficient (r(2)) in dogs between manometry and applanation tonometers Tono-Pen XL (R) and Perkins (R) were, respectively, 0.896 and 0.981 and in cats were 0.905 and 0.988. The mean IOP values in conscious dogs with Tono-Pen XL (R) and Perkins (R) were, respectively, 17.5 +/- 3.7 mmHg (10.0-25.0 mmHg) and 15.3 +/- 2.1 mmHg (12.0-19.8 mmHg) and in conscious cats were 16.8 +/- 3.6 mmHg (10.5-24.5 mmHg) and 15.5 +/- 1.3 mmHg (13.0-18.5 mmHg). Conclusion There was a strong correlation between the IOP values obtained by direct ocular manometry and the Tono-Pen XL (R) and Perkins (R) tonometers in dogs and cats. There was no statistically significant difference between the mean IOP obtained with both tonometers in conscious animals, there was, however, a difference between the minimum and mainly in the maximum values that were on average 5-6 mmHg higher with Tono-Pen XL (R) than those measured with Perkins (R), which justifies a table of normal values differentiated for each tonometer.
机译:目的比较在眼底镜正常的狗和猫中两种压平眼压计Tono-Pen XL(R)和Perkins(R)的准确性。动物评价了25只意识和健康的猫狗的两只眼睛。死后立即用五只狗和猫的两只眼睛作为死后研究的对照。程序在有意识的动物中,使用局部麻醉对眼压计进行眼压测定,两个眼压计均使用0.5%代卡因他明滴眼液,对于Perkins眼压计,使用1%荧光素眼药水。验尸研究中眼压(IOP)的读数是通过Tono-Pen XL(R)和Perkins(R)使用测压法和眼压计获得的。结果测压和眼压眼压计Tono-Pen XL(R)和Perkins(R)在狗中的相关系数(r(2))分别为0.896和0.981,在猫中为0.905和0.988。使用Tono-Pen XL(R)和Perkins(R)的清醒狗的平均IOP值分别为17.5 +/- 3.7 mmHg(10.0-25.0 mmHg)和15.3 +/- 2.1 mmHg(12.0-19.8 mmHg)和在有意识的猫中,其体重分别为16.8 +/- 3.6 mmHg(10.5-24.5 mmHg)和15.5 +/- 1.3 mmHg(13.0-18.5 mmHg)。结论通过直接眼压测量获得的IOP值与狗和猫的Tono-Pen XL(R)和Perkins(R)眼压计之间存在很强的相关性。在有意识的动物中,两种眼压计获得的平均眼压之间无统计学显着性差异,但Tono-Pen XL(R的最小值和主要最大值之间的差异平均高出5-6 mmHg(R ),而不是用珀金斯(R)测量的值,这证明了针对每个眼压计的正常值表的合理性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号