首页> 外文期刊>Veterinary Ophthalmology >Tono-Pen XL calibration curves for cats, cows and sheep
【24h】

Tono-Pen XL calibration curves for cats, cows and sheep

机译:猫,牛和羊的Tono-Pen XL校准曲线

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The objective of this study was to provide calibration curves for correcting intraocular pressure (IOP) measurements obtained using the Tono-Pen XL tonometer in cats, cows and sheep. Twelve eyes from 9 cats, 13 eyes from 7 cows, 10 eyes from 5 sheep were used. The anterior chamber of the eye was cannulated in vivo, in situ (immediately post mortem) or ex vivo with a fine needle and IOP was varied from 10 to 90 mmHg in steps of 10 mmHg by adjusting the height of a saline reservoir connected to the needle. For each pressure setting, several readings of IOP were made using the tonometer. The relationship between Tono-Pen reading and manometer setting was linear over the full range of measurement. However, the slope of the data regression line deviated significantly from 1 and indicated that the instrument systematically underestimated IOP. For cats the average slope was 0.62 and for cows and sheep it was 0.72 and 0.69, respectively. For the latter animals, the regression line also had a nonzero intercept of approximately 4.5 mmHg. Similar results were obtained from in vivo and ex vivo eyes and with different Tono-Pen XL tonometers. Although developed for use on humans, the Tono-Pen XL can provide reproducible and accurate measurement of IOP in cats, cows and sheep when suitably calibrated by manometry. The calibration curves provided here, and by implication those reported for other animals using this tonometer, differ in slope from those measured with earlier models of the Tono-Pen. The reproducibility of the curves we obtained implies that they can be used to correct IOP readings from the Tono-Pen XL when manometry is not possible.
机译:这项研究的目的是提供校正曲线,以校正使用Tono-Pen XL眼压计在猫,牛和羊中获得的眼内压(IOP)测量值。使用了来自9只猫的十二只眼,来自七头牛的13只眼,来自五只羊的10只眼。用细针在体内,原位(死后立即进行)或离体插管眼前房,并通过调节与食管相连的生理盐水的高度,将眼压从10mmHg改变为10mmHg至90mmHg。针。对于每个压力设置,使用眼压计读取IOP的多个读数。 Tono-Pen读数与压力计设置之间的关系在整个测量范围内都是线性的。但是,数据回归线的斜率显着偏离1,表明该仪器系统地低估了IOP。猫的平均斜率是0.62,牛和羊的平均斜率分别是0.72和0.69。对于后者的动物,回归线也具有约4.5 mmHg的非零截距。从体内和离体眼睛以及使用不同的Tono-Pen XL眼压计可获得类似的结果。尽管为人类使用而开发,但Tono-Pen XL在通过测压法进行适当校准后,可提供可重复且准确的猫,牛和羊眼内压测量。此处提供的校准曲线以及暗示使用该眼压计为其他动物报告的校准曲线的斜率与使用Tono-Pen早期模型测得的斜率不同。我们获得的曲线的可再现性意味着,当不可能进行测压时,可以将其用于校正Tono-Pen XL的IOP读数。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号