A monolithic x-ray focusing lens can focus divergent x-rays from a conventional x-ray tube into 110 mu m microbeam. A gain factor of more than 1200 compared with an aperture without the lens has been achieved. Using a monolithic x-ray focusing lens combined with aperture optics, a 50 mu m microbeam with sufficient intensity has been obtained, The detection limit for Mn by means of the x-rag microbeam source is approximately 1 x 10(-12) g when a 1.5 W x-ray tube power and a 1000 s measurement time are employed. (C) 1997 John Wiley & Sons, Ltd. [References: 14]
机译:单片X射线聚焦透镜可以将来自常规X射线管的发散X射线聚焦到110微米的微束中。与不带镜头的光圈相比,增益系数已超过1200。使用单片x射线聚焦透镜结合光阑光学元件,可获得强度足够的50μm微束。当x-rag微束源检测到Mn时,Mn的检出限约为1 x 10(-12)g使用1.5 W X射线管功率和1000 s测量时间。 (C)1997 John Wiley&Sons,Ltd. [参考:14]