Determination of Cr(III) and Cr(VI) in water by wavelength-dispersive X-ray fluorescence spectrometry after preconcentration with an ion-exchange resin disk

机译:离子交换树脂盘预浓缩后,通过波长色散X射线荧光光谱法测定水中的Cr(III)和Cr(VI)

获取原文
获取原文并翻译 | 示例
           

摘要

A rapid and simple method using an ion-exchange resin disk combined with wavelength-dispersive X-ray fluorescence (WDXRF) spectrometry was developed for the determination of Cr(III) and Cr(VI) in water. A 100-ml water sample was first adjusted to pH 3 with nitric acid and then passed through an anion-exchange resin disk placed on top of a cation-exchange resin disk at a flow rate of 1 ml min~(-1) to separate Cr(III) and Cr(VI). Anionic Cr(VI) was preconcentrated on the upper anion-exchange resin disk, whereas cationic Cr(III) was preconcentrated on the lower cation-exchange resin disk. Each ion-exchange resin disk was dried at 100 degrees C for 30 min in an electric oven and coated with a commercially available laminate film. The specimens were measured using a WDXRF spectrometer. The calibration curves of Cr(III) and Cr(VI) showed good linearity in the range 1-10μg. The detection limits corresponding to three times the standard deviation (n = 5) of blank values were 0.17μg for Cr(III) and 0.16 mu g for Cr(VI). If a 1-I water sample is used, these limits would be 0.17 and 0.16 μg l~(-1), respectively. A spike test for 50 μgl~(-1) Cr(III) and Cr(VI) in tap water and river water showed quantitative recoveries (94-114%), although this was not observed for mineral drinking water owing to the overlap of V K beta with Cr K alpha. The recovery after overlap correction was satisfactory (115%).
机译:建立了一种使用离子交换树脂圆盘结合波长色散X射线荧光(WDXRF)光谱法的快速简便的方法,用于测定水中的Cr(III)和Cr(VI)。首先用硝酸将100毫升水样品的pH值调节至3,然后以1毫升min〜(-1)的流速通过放置在阳离子交换树脂盘顶部的阴离子交换树脂盘以分离Cr(III)和Cr(VI)。阴离子Cr(VI)预浓缩在上层阴离子交换树脂盘上,而阳离子Cr(III)预浓缩在下层阳离子交换树脂盘上。将每个离子交换树脂盘在电烤箱中在100摄氏度下干燥30分钟,并涂以市售的层压膜。使用WDXRF光谱仪测量样品。 Cr(III)和Cr(VI)的校准曲线在1-10μg范围内显示出良好的线性。对应于空白值标准偏差(n = 5)的三倍的检出限为Cr(III)为0.17μg,Cr(VI)为0.16μg。如果使用1-I水样品,则这些限值分别为0.17和0.16μgl〜(-1)。对自来水和河水中50μgl〜(-1)Cr(III)和Cr(VI)的加标测试显示了定量回收率(94-114%),尽管由于矿泉水的重叠,未观察到矿物质饮用水的回收率VK beta和Cr K alpha。重叠校正后的恢复令人满意(115%)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号