Automated preparation of platelet concentrates from pooled buffy coats: in vitro studies and experiences with the OrbiSac system.

机译:从合并的血沉棕黄层中自动制备血小板浓缩物:OrbiSac系统的体外研究和经验。

获取原文
获取原文并翻译 | 示例

摘要

BACKGROUND: The aim was to evaluate platelet concentrates (PCs) prepared by the automated OrbiSac system, from pooled buffy coats (BCs) stored in a platelet (PLT) additive solution. STUDY DESIGN AND METHODS: Experiment 1 was a paired in vitro study of PCs (from six BCs), prepared by automated and manual procedures. Experiments 2 and 3 evaluated PCs from OrbiSac (from six BCs); Experiment 3 included selection of BCs based on donor data. Experiment 4 was a paired in vitro study of PCs (from six BCs) with an integrated white blood cell (WBC) filter and two different storage containers. Experiment 5 evaluated PCs (from six BCs) from the OrbiSac with an integrated WBC filter. Experiment 6 was similar to Experiment 5 with computer-selected pools of 5 BCs. The in vitro studies evaluated the effects of 7-day storage of PLTs regarding PLT metabolism and disintegration. RESULTS: Experiments 1 and 4 had similar in vitro results. In Experiment 2, PLT content was 370 x 10(9) +/- 70 x 10(9) per PC and recovery from BCs was 76 +/- 6 percent. In Experiment 3, the PLT content was 380 x 10(9) +/- 50 x 10(9) per PC and variation was reduced compared with randomly pooled BCs. In Experiment 5, increased PLT content was found (420 x 10(9) +/- 70 x 10(9) per PC and recovery from BCs of 80 +/- 5%). In Experiment 6, five rather than six BCs gave 340 x 10(9) +/- 60 x 10(9) PLTs per PC and recovery was 79 +/- 5 percent. CONCLUSION: These in vitro studies suggest that the OrbiSac technique is equivalent to the standard manual method regarding the PLT in vitro characteristics during storage for 7 days. The results of standardizing the PLT count in PCs by selecting the BCs pools on the basis of the blood donor PLT concentration were encouraging.
机译:背景:目的是评估自动OrbiSac系统从储存在血小板(PLT)添加剂溶液中的血沉棕黄层(BC)中制备的血小板浓缩物(PC)。研究设计和方法:实验1是通过自动和手动程序对PC(来自六个BC)进行的配对体外研究。实验2和3评估了OrbiSac的PC(来自六个BC);实验3包括根据供体数据选择BC。实验4是对PC(来自六个BC),带有集成白细胞(WBC)过滤器和两个不同存储容器的体外配对研究。实验5评估了带有集成WBC滤波器的OrbiSac的PC(来自六个BC)。实验6与实验5类似,但计算机选择的5个BC池。体外研究评估了7天储存PLT对PLT代谢和分解的影响。结果:实验1和4具有相似的体外结果。在实验2中,每台PC的PLT含量为370 x 10(9)+/- 70 x 10(9),从BC中的回收率为76 +/- 6%。在实验3中,每台PC的PLT含量为380 x 10(9)+/- 50 x 10(9),并且与随机合并的BC相比,变化减少了。在实验5中,发现PLT含量增加了(每台PC 420 x 10(9)+/- 70 x 10(9),从BC中恢复80 +/- 5%)。在实验6中,五台而不是六台BC提供了每台PC 340 x 10(9)+/- 60 x 10(9)的PLT,回收率为79 +/- 5%。结论:这些体外研究表明,OrbiSac技术等同于7天储存期间PLT体外特性的标准手动方法。通过根据献血者PLT浓度选择BC池来标准化PC中PLT计数的结果令人鼓舞。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号