Evaluation of polyethylene terephthalate for ABO and Rh typing and alloantibody screening.

机译:评估聚对苯二甲酸乙二酯的ABO和Rh类型以及同种抗体筛选。

获取原文
获取原文并翻译 | 示例

摘要

BACKGROUND: For many years, hospitals and laboratories have used evacuated glass tubes for blood collection. To improve the safety of blood collection, plastic polyethylene terephthalate (PET) tubes (Vacutainer PLUS, Becton Dickinson) have been developed. The objectives of this study were to compare the accuracy of ABO grouping, Rh typing, and antibody screening of blood samples collected in plastic tubes with that in glass tubes and to determine if refrigerated blood samples collected in plastic tubes remained stable over a 28-day period. STUDY DESIGN AND METHODS: Samples were collected from 121 volunteers, at least 30 from each of the A, B, O, and AB blood groups, in four types of Vacutainer tubes: silica-coated plastic, K(2) EDTA plastic, uncoated glass, and K(2) EDTA glass. Samples from each tube were tested for ABO group and Rh type by use of the microtyping gel identification card system and the tube method. A three-cell antibody screen was performed by the microtyping gel card technique with a monospecific IgG reagent. Initial samples were tested within 3 hours of collection. Refrigerated samples were retested for ABO and Rh type and antibody screening 1, 2, 21, and 28 days later. Agreement between test results was determined by using Cohen's Kappa statistic. RESULTS: Complete agreement was observed between the ABO and Rh typing results in samples drawn into glass and plastic tubes of both the EDTA and nonanticoagulated type (kappa = 1.0). In retesting, there were no examples of a change in ABO or Rh type over the 28-day study period. Only two alloantibodies (1.7%) were identified in the 121 samples, and no difference was observed in alloantibody expression in either plastic or glass Vacutainer tubes over the 28-day study period. CONCLUSION: Samples collected into the PET serum or EDTA tubes provided accurate ABO and Rh typing results that remained consistent over a 28-day period. Samples collected in these tubes also appeared to enable accurate alloantibody identification. However, the number of alloantibodies identified in this study was small, and this result should be confirmed in a larger series.
机译:背景:多年来,医院和实验室已使用真空玻璃管进行血液采集。为了提高采血的安全性,已经开发了塑料聚对苯二甲酸乙二醇酯(PET)管(Vacutainer PLUS,Becton Dickinson)。这项研究的目的是比较ABO分组,Rh分型和在塑料管中收集的血液样本与玻璃管中的血液样本进行抗体筛查的准确性,并确定在28天中塑料管中收集的冷藏血液样本是否保持稳定期。研究设计和方法:从121种志愿者中收集了样本,其中至少30种分别来自A,B,O和AB血型,分别使用四种类型的Vacutainer管:硅胶涂层塑料,K(2)EDTA塑料,未涂层玻璃和K(2)EDTA玻璃。使用微量分型凝胶识别卡系统和试管方法,测试每个试管的样品的ABO组和Rh类型。通过微分型凝胶卡技术用单特异性IgG试剂进行三细胞抗体筛选。初始样品在收集后的3小时内进行了测试。重新测试冷藏样品的ABO和Rh类型,并在1、2、21和28天后进行抗体筛选。使用Cohen的Kappa统计量确定测试结果之间的一致性。结果:将样品吸入EDTA和非抗凝型(kappa = 1.0)的玻璃和塑料管中的样品,ABO和Rh分型结果完全一致。在重新测试中,没有28天研究期间ABO或Rh类型改变的例子。在121天的样本中,仅鉴定出两种同种抗体(1.7%),在28天的研究期内,在塑料或玻璃Vacutainer管中的同种抗体表达均未观察到差异。结论:收集到PET血清或EDTA试管中的样品提供了准确的ABO和Rh分型结果,在28天内保持一致。从这些试管中收集的样品也似乎可以准确鉴定同种抗体。但是,本研究中鉴定出的同种抗体数量很少,应在更大系列中证实这一结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号