Analytical and clinical validation of a new point-of-care testing system for determination of D-Dimer in human blood.

机译:用于确定人血中D-二聚体的新型即时检验系统的分析和临床验证。

获取原文
获取原文并翻译 | 示例
           

摘要

D-Dimer testing is used for exclusion of deep venous thrombosis (DVT). AQT90 FLEX D-Dimer (AQT D-Dimer) is a novel time-resolved fluorescence based point-of-care test for quantification of D-Dimer in whole blood or plasma. Presently we have determined the analytical and clinical performance of AQT D-Dimer and compared it with four routine D-Dimer assays. The within-run CV of AQT D-Dimer was 3.8-7.2% and the between-run CV was 5.7- 9.7%. Excellent agreement was found between the D-Dimer concentrations recorded in citrate-, heparin- and EDTA stabilised blood. The plasma concentration of D-Dimer was determined with AQT D-Dimer, AxSYM, Biopool Auto-Dimer, STA-Liatest and Vidas New in 170 consecutive patients suspected for DVT. Phlebograms were positive in 64 patients (22 distal, 42 proximal). ROC-curves (ROC), the negative and positive predictive values (NPV, PPV), the sensitivity and specificity of the tests were compared. The area under ROC was comparable for all tests. NPV for all DVT was 87-88%, the sensitivity was 88-92% and the PPV was 45-55%. For proximal DVT the NPV and sensitivity were 100% for all tests, whereas the PPV was 37-48%. For distal DVT we obtained a NPV of 87-88%. The sensitivity was 64-77%, the PPV was 19-24% whereas a specificity of 32-58% was observed. The AQT D-Dimer demonstrates excellent analytical and diagnostic performance. The test is rapidly performed and the measuring range of the assay is wide. The NPV, PPV, specificity, sensitivity and AUC of AQT D-Dimer for both proximal and distal DVT are comparable to routine D-Dimer assays.
机译:D-Dimer检测用于排除深静脉血栓形成(DVT)。 AQT90 FLEX D-Dimer(AQT D-Dimer)是一种基于时间分辨荧光的新型即时检验,用于定量全血或血浆中的D-Dimer。目前,我们已经确定了AQT D-Dimer的分析和临床性能,并将其与四种常规D-Dimer分析进行了比较。 AQT D-Dimer的运行内CV为3.8-7.2%,运行间CV为5.7-9.7%。在柠檬酸盐,肝素和EDTA稳定的血液中记录的D-二聚体浓度之间发现了极好的一致性。使用AQT D-二聚体,AxSYM,Biopool自动二聚体,STA-Liatest和Vidas New在170名怀疑患有DVT的连续患者中测定D-二聚体的血浆浓度。静脉造影阳性的患者有64例(远端22例,近端42例)。比较了ROC曲线(ROC),阴性和阳性预测值(NPV,PPV),检测的敏感性和特异性。在所有测试中,ROC下的面积均相当。所有DVT的NPV为87-88%,灵敏度为88-92%,PPV为45-55%。对于近端DVT,所有测试的NPV和灵敏度均为100%,而PPV为37-48%。对于远端DVT,我们获得了87-8%的NPV。敏感性为64-77%,PPV为19-24%,而观察到的特异性为32-58%。 AQT D-Dimer具有出色的分析和诊断性能。测试迅速进行,测定的测量范围广。近端和远端DVT的AQT D-Dimer的NPV,PPV,特异性,敏感性和AUC与常规D-Dimer分析相当。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号