Evaluation of two automated methods for measurement of the ristocetin cofactor activity of von Willebrand factor.

机译:评估两种自动方法来测量von Willebrand因子的ristocetin辅因子活性。

获取原文
获取原文并翻译 | 示例

摘要

We have evaluated two automated methods for measuring ristocetin cofactor activity (VWF:RCo) on an automated coagulation analyzer (STAR, Diagnostica Stago Inc., Parsippany, NJ). A modification of the BC von Willebrand Reagent method (Dade Behring Inc., Newark, Delaware) was compared with a standard method using a platelet aggregometer, and an in-house automated method was developed using commercially available components. The STAR measures the change in optical density (OD) caused by agglutination of platelets. Change in OD was linear for plasma dilutions between .25 and .04 U/mL. Coefficient of variation compared favorably with that of the standard method. In 123 normal women, the correlation coefficient (r) = .905. In menorrhagia cases, r = .901. In 20 VWD patients, r = .93. The CDC in-house method was compared to the BC Reagent method in 79 additional menorrhagia cases, with r = .94. The automated methods produced results equivalent or superior to those of traditional methods of measuring VWF:RCo.
机译:我们已经评估了两种在自动凝血分析仪(STAR,Diagnostica Stago Inc.,Parsippany,NJ)上测量瑞斯托素辅助因子活性的自动方法(VWF:RCo)。将BC von Willebrand试剂方法的一种改进方法(Dade Behring Inc.,特拉华州纽瓦克)与使用血小板凝集仪的标准方法进行了比较,并使用市售组件开发了内部自动方法。 STAR测量由血小板凝集引起的光密度(OD)的变化。当血浆稀释度介于0.25至0.04 U / mL之间时,OD值呈线性变化。变异系数比标准方法优越。在123名正常女性中,相关系数(r)= .905。在月经过多的情况下,r = .901。在20名VWD患者中,r = 0.93。在另外79例月经过多病例中,将CDC内部方法与BC试剂方法进行了比较,r = 0.94。自动化方法产生的结果与传统的VWF:RCo测量方法相当或更好。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号