This study was conducted to compare the cross-reactivity of two commercially available carbamazepine (CBZ) immunoassays (PETINIA and EMIT 2000) with carbamazepine-10,11-epoxide (CBZ-E), the active metabolite of CBZ. Oxcarbazepine (OCBZ) and its main metabolite 10-hydroxy-carbazepine (HCBZ) have a chemical structure closely related to that of CBZ. The cross-reactivities of these two drugs were also investigated. In the first part of the study, Lyphocheck blank human serum and Chemonitor quality controls (containing CBZ without CBZ-E) were spiked with variable amounts of CBZ-E. The apparent CBZ levels were measured by PETINIA and EMIT 2000 methods. The interference from OCBZ and HCBZ was directly assessed by measuring the apparent CBZ levels in Chromsystems Trileptal quality controls (containing OCBZ and HCBZ). In the second part of the study, the CBZ levels of serum samples from 49 patients, including 2 patients with massive CBZ ingestion, were measured by immunoassays and compared with a high-pressure liquid chromatography (HPLC) reference technique allowing the simultaneous measurement of CBZ and CBZ-E. The antibody used in the PETINIA assay cross-reacts (about 90%) with CBZ-E. In one case of CBZ poisoning (CBZ and CBZ-E levels measured by HPLC were 26.2 and 18.2 mg/L, respectively), CBZ level measured by PETINIA was falsely elevated (42.5 mg/L). In contrast, the specificity of EMIT 2000 was satisfactory (29.5 mg/L). The two immunoassays tested showed low cross-reactivity with OCBZ and HCBZ. In conclusion, it appears that the CBZ-E metabolite present in samples can falsely increase CBZ levels measured by the PETINIA assay.
机译:进行这项研究的目的是比较两种市售的卡马西平(CBZ)免疫测定法(PETINIA和EMIT 2000)与卡马西平-10,11-环氧化物(CBZ-E)(CBZ的活性代谢物)的交叉反应性。奥卡西平(OCBZ)及其主要代谢物10-羟基-卡巴西平(HCBZ)的化学结构与CBZ密切相关。还研究了这两种药物的交叉反应性。在研究的第一部分中,将Lyphocheck空白人血清和Chemonitor质量对照品(包含不含CBZ-E的CBZ)掺入不同量的CBZ-E。通过PETINIA和EMIT 2000方法测量了表观CBZ水平。通过测量Chromsystems Trileptal质量对照品(包含OCBZ和HCBZ)中的表观CBZ水平,可以直接评估OCBZ和HCBZ的干扰。在研究的第二部分中,通过免疫测定法测量了49名患者(包括2名大量摄入CBZ的患者)的血清样品的CBZ水平,并与允许同时测量CBZ的高压液相色谱(HPLC)参考技术进行了比较和CBZ-E。 PETINIA分析中使用的抗体与CBZ-E发生交叉反应(约90%)。在CBZ中毒的一例中(通过HPLC测量的CBZ和CBZ-E水平分别为26.2和18.2 mg / L),通过PETINIA测量的CBZ水平错误地升高(42.5 mg / L)。相反,EMIT 2000的特异性令人满意(29.5 mg / L)。所测试的两种免疫测定显示与OCBZ和HCBZ的交叉反应性低。总之,看来样品中存在的CBZ-E代谢物可能会错误地增加通过PETINIA分析测得的CBZ水平。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号