首页> 外文期刊>The Veterinary Journal >Evaluation of the veterinary application of a point-of-care device measuring white blood cell counts
【24h】

Evaluation of the veterinary application of a point-of-care device measuring white blood cell counts

机译:兽医测量白血球计数的现场护理设备的评估

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A point-of-care device (POCD) for measuring total white blood cell count was evaluated for feline, canine, equine and bovine blood samples collected into EDTA. Mean biases were ?9.2% (range, ?12% to ?6.3%) for feline samples, 20.2% (range, 15.3–25.1%) for canine samples, ?7.1% (range, ?8.3% to ?5.9%) for equine samples, and 0.7% (range, ?1.1% to 2.5%) for bovine samples. The results were influenced by the presence of nucleated red blood cells. The POCD provided precise, reliable data for feline, equine and bovine samples but the values obtained for the canine counts were overestimations.
机译:对用于测量总白细胞计数的即时医疗设备(POCD)评估了收集到EDTA中的猫,犬,马和牛血样。猫样品的平均偏差为?9.2%(范围为?12%至?6.3%),犬样品的平均偏差为20.2%(范围为15.3-25.1%),样品为?7.1%(范围为?8.3%至?5.9%)。马样品,牛样品为0.7%(范围从1.1%到2.5%)。结果受有核红细胞的存在影响。 POCD为猫,马和牛样品提供了准确,可靠的数据,但犬计数获得的值却被高估了。