In order to provide objective measures of anaesthetic depth for periods without clinical signs (e.g., cardiopulmonary bypass [CPB]), the correlation between the electroencephalographic Narcotrend index (NI), 95% spectral edge frequency (SEF95), median power frequency (MPF) and clinical stages of anaesthesia was investigated in 16 juvenile sheep. Data were recorded during recovery from anaesthesia for pulmonary or aortic valve replacement. A significant (P < 0.05) negative correlation was found between clinical stages of anaesthesia and NI (r(s) = -0.534) and SEF95 (r(s) = -0.543). No significant correlation existed between anaesthetic stages and MPF (r(s) = -0.292, P > 0.05). The sensitivity of NI, SEF95 and MPF to assure an adequate level of anaesthesia was 71.43%, 66.67% and 44.44%, respectively, while the specificity of the descriptors ranged between 97.44 and 92.31%. No significant age-related effect on EEG data and stages of anaesthesia was detected when data from sheep between 4 and 6 months of age were compared with data obtained from 8- to 12-month old sheep. In conclusion, NI seems to be the most appropriate EEG descriptor to assure adequate depth of anaesthesia in juvenile, isoflurane-anaesthetized sheep
机译:为了提供无临床症状(例如,体外循环[CPB])期间麻醉深度的客观测量,脑电图Narcotrend指数(NI),95%频谱边缘频率(SEF95),中位功率频率(MPF)之间的相关性并在16只小绵羊中调查了麻醉的临床阶段。从麻醉恢复期间记录数据以进行肺或主动脉瓣置换。在麻醉与NI(r(s)= -0.534)和SEF95(r(s)= -0.543)的临床阶段之间发现显着(P <0.05)负相关。麻醉阶段与MPF之间无显着相关性(r = -0.292,P> 0.05)。 NI,SEF95和MPF确保足够麻醉的敏感性分别为71.43%,66.67%和44.44%,而描述符的特异性在97.44和92.31%之间。将4到6个月大的绵羊的数据与8到12个月大的绵羊的数据进行比较,未发现对EEG数据和麻醉阶段有明显的年龄相关影响。总而言之,NI似乎是确保幼年异氟烷麻醉的绵羊足够麻醉深度的最合适的EEG描述子

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号