Use of the CDI blood parameter monitoring system 500 for continuous blood gas measurement during extracorporeal membrane oxygenation simulation.

机译:CDI血液参数监测系统500在体外膜氧合模拟期间用于连续血气测量的用途。

获取原文
获取原文并翻译 | 示例
           

摘要

The Oximetrix III Opticath (Abbott Critical Care Systems) is used for continuous measurement of venous saturation in a variety of applications, including extracorporeal membrane oxygenation (ECMO), despite clinical reports that have presented data showing poor accuracy of these devices. The CDI Blood Parameter Monitoring System 500 (Terumo) is an inline blood gas monitoring tool commonly used during cardiopulmonary bypass procedures to continuously assess oxygen saturation, blood gases, potassium, and bicarbonate. The purpose of this experiment was to compare the Opticath and the CDI 500 in trending venous blood saturation during a simulation of ECMO. An ECMO simulation circuit consisting of a silicone rubber membrane oxygenator and a stainless steel heat exchanger was constructed, and a standard venous reservoir bag was used to represent the patient. The CDI and the Opticath were incorporated side by side into a shunt that originated just before the oxygenator and returned flow to the venous line. Thecircuit was primed with fresh porcine blood and conditioned with the addition of CO2 to simulate typical venous blood under ECMO conditions. After an initial calibration procedure, samples were drawn and analyzed by an AVL Opti CCA (Roche/Osmetech) every 4-8 hours for a period of 7 days, with calibration of each device at sample intervals. The data were plotted, and a least squares regression line was calculated. The average error for venous saturation of the CDI and Opticath after 72 hours was 3.86 and 9.51 respectively. At 168 hours, error for the CDI was 8.37, and the Opticath had an error of 14.78. A correlation analysis of the CDI and AVL CCA analyzer yielded a correlation coefficient of r = .88 at 72 hours and r = .84 at 168 hours. Correlation between the Opticath and the AVL CCA yielded a correlation coefficient of r = .77 at 72 hours and r = .55 at 168 hours. Based on these findings, the CDI 500 is an effective tool for monitoring venous blood saturation under simulated conditions of ECMO. Keywords: CDI 500, Opticath, extracorporeal membrane oxygenation, venous saturation.
机译:尽管临床报告提供的数据表明这些设备的准确性较差,但Oximetrix III Opticath(Abbott重症监护系统)在包括体外膜氧合(ECMO)在内的各种应用中用于连续测量静脉饱和度。 CDI血液参数监测系统500(Terumo)是一种在线血气监测工具,通常在心肺旁路手术过程中使用,以连续评估氧饱和度,血气,钾和碳酸氢盐。本实验的目的是在ECMO模拟期间比较Opticath和CDI 500在趋势静脉血饱和度方面的对比。建立了由硅橡胶膜充氧器和不锈钢热交换器组成的ECMO模拟电路,并使用标准静脉储液袋来代表患者。 CDI和Opticath并排并入一个分流器,该分流器刚好在充氧器之前产生,并返回到静脉管线。电路用新鲜的猪血灌注,并加入二氧化碳调节,以模拟ECMO条件下的典型静脉血。进行初始校准程序后,每4-8小时通过AVL Opti CCA(Roche / Osmetech)抽取样品并进行分析,为期7天,并以样品间隔对每个设备进行校准。绘制数据,并计算最小二乘回归线。 72小时后CDI和Opticath的静脉饱和度的平均误差分别为3.86和9.51。在168小时时,CDI的错误为8.37,而Opticath的错误为14.78。 CDI和AVL CCA分析仪的相关分析得出的相关系数在72小时时为r = 0.88,在168小时时为r = 0.84。 Opticath和AVL CCA之间的相关性得出的相关系数在72小时时为r = 0.77,在168小时时为r = 0.55。基于这些发现,CDI 500是在ECMO模拟条件下监测静脉血饱和度的有效工具。关键字:CDI 500,Opticath,体外膜充氧,静脉饱和度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号