The aim of the present study was to report our additional experience with transcatheter closure of the patent ductus arteriosus in 65 consecutive patients using the new Amplatzer duct occluder. The median patient age was 3.6 years (range 0.2to 12), and the median weight was 10.5 kg (range 4 to 38). The device was a modified Amplatzer duct occluder made of fabric-free fine Nitinol wire net in to 2 very low profile disks with an articulated connecting waist. It is delivered through a 4Fr to 5Fr delivery sheath. The device was permanently implanted in 62 of 65 patients. The mean patent ductus arteriosus diameter (at the pulmonary end) was 3.6 +/- 1.3 mm (range 0.5 to 5.5). The mean device diameter (waist diameter) was 4.2 +/- 1.5 mm (range 3 to 6). Complete echocardiographic closure of the PDA at 1 month follow-up was observed in 61 (98%) of 62 patients. Immediately after the procedure, mild left pulmonary stenosis (peak pressure gradient of 8, 10, and 12 mm Hg) in 3 of 63 patients. Device embolization in 1 patient was the main complication of the procedure. No other complications were observed. In conclusion, catheter closure using the Amplatzer duct occluder II is an effective and safe therapy for most patients with patent ductus arteriosus. Additional studies are required to document its efficacy, safety, and long-term results in a larger patient population.
机译:本研究的目的是报告使用新的Amplatzer导管封堵器连续65例患者经导管闭合动脉导管未闭的附加经验。患者的中位年龄为3.6岁(范围为0.2至12),体重的中位数为10.5 kg(范围为4至38)。该设备是一种改良的Amplatzer风道阻塞器,由无织物的细镍钛诺金属丝网制成,分成2个非常低矮的圆盘,带有铰接的连接腰部。它通过4Fr至5Fr输送护套输送。该装置已永久植入65位患者中的62位。平均动脉导管未闭直径(在肺端)为3.6 +/- 1.3毫米(0.5到5.5)。装置的平均直径(腰部直径)为4.2 +/- 1.5毫米(范围3至6)。 62例患者中有61例(98%)在1个月的随访中完全关闭了PDA的超声心动图检查。手术后立即在63例患者中的3例中出现轻度左肺狭窄(峰值压力梯度为8、10和12 mm Hg)。 1例患者的设备栓塞是该过程的主要并发症。没有观察到其他并发症。总之,对于大多数动脉导管未闭的患者,使用Amplatzer导管封堵器II进行导管闭合是一种有效且安全的疗法。需要更多的研究来证明其在更多患者中的功效,安全性和长期结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号