首页> 外文期刊>The Journal of the Acoustical Society of America >Frequency discrimination and speech recognition by patients who use the Ineraid and continuous interleaved sampling cochlear-implant signal processors
【24h】

Frequency discrimination and speech recognition by patients who use the Ineraid and continuous interleaved sampling cochlear-implant signal processors

机译:使用Ineraid和连续交错采样的人工耳蜗信号处理器的患者的频率识别和语音识别

获取原文
获取原文并翻译 | 示例
           

摘要

Patients who use the Ineraid cochlear implant were tested in four experiments with materials which assessed frequency discrimination and speech understanding. In each experiment both frequency discrimination and speech recognition varied among patients. Correlations between the two measures were significant and ranged from 0.60 to 0.83. Most generally, frequency discrimination was better in the frequency domain of F1 than in the domain of F2. In experiment 5, both the Ineraid signal processing strategy and a continuous interleaved sampling (CIS) strategy were implemented for a single patient. The CIS strategy improved frequency discrimination in the domain of F2 and improved speech understanding.
机译:使用Ineraid人工耳蜗的患者在四个实验中接受了评估频率歧视和语音理解的材料的测试。在每个实验中,患者的频率识别和语音识别都不同。两项测量之间的相关性很显着,范围为0.60至0.83。最普遍的是,F1的频域比F2的频域更好。在实验5中,为单个患者实施了Ineraid信号处理策略和连续交错采样(CIS)策略。 CIS策略改善了F2域中的频率歧视并改善了语音理解。