首页> 外文期刊>The Journal of Thoracic and Cardiovascular Surgery >Forty-one-month follow-up of the Symmetry aortic connector system for proximal venous anastomosis.
【24h】

Forty-one-month follow-up of the Symmetry aortic connector system for proximal venous anastomosis.

机译:Symmetry主动脉连接器系统的41个月随访,用于近端静脉吻合。

获取原文
获取原文并翻译 | 示例

摘要

OBJECTIVE: Results of short- and midterm follow-up studies of the patency rate of the Symmetry aortic connector systems (St Jude Medical, Inc, Minneapolis, Minn) are controversial. Long-term follow-up studies are still lacking (so far, the longest mean follow-up period was 19 months). The aim of our study was (1) to evaluate the patency rate of this device over a longer time-period and (2) to analyze risk factors for graft occlusion. METHODS: Between November 2000 and July 2003, 76 Symmetry aortic connector systems were implanted in 42 patients. At follow-up, 24 patients with 44 mechanical connectors were studied with 64-slice cardiac computed tomography. Eight patients had died previously, 6 patients refused to undergo a computed tomographic scan, and 4 patients had to be excluded because of impaired renal function. RESULTS: From a total of 44 mechanical connectors studied, 24 (55%) were occluded, 20 (45%; confidence intervals 31%-61%) were patent, and 7 of these grafts showed stenosis in the area of the connector. Mean follow-up was 41 +/- 10 months (18-52 months). Sex, age, left main stenosis, hyperlipidemia, hypertension, renal failure, target vessel, stenosis of the target vessel, diameter of the target vessel, type of surgical intervention, diabetes, ejection fraction, postoperative anticoagulation regimen, and the connector size showed no significant influence on the bypass graft patency (P > .05). The bypass graft flow was recognized to be the only risk factor for bypass graft occlusion (P = .0256). CONCLUSION: Midterm follow-up data show a high number of occluded Symmetry aortic connector system vein grafts. On the basis of these observations, the use of the connector was abandoned at our institution.
机译:目的:对称性主动脉连接器系统(St Jude Medical,Inc,明尼阿波利斯,明尼苏达州)的通畅率的短期和中期随访研究结果存在争议。尚缺乏长期随访研究(迄今为止,最长的平均随访期为19个月)。我们的研究目的是(1)评估该装置在较长时间内的通畅率,以及(2)分析移植物闭塞的危险因素。方法:2000年11月至2003年7月,共向42例患者植入了76个对称主动脉连接器系统。随访时,对64例心脏机械连接的24例患者进行了64层心脏计算机断层扫描。先前有8例患者死亡,有6例患者拒绝进行计算机断层扫描,而4例患者由于肾功能受损而被排除在外。结果:总共研究了44个机械连接器,其中有24个(55%)被阻塞,有20个(45%;置信区间为31%-61%)被授予专利权,其中7个移植物在连接器区域显示狭窄。平均随访时间为41 +/- 10个月(18-52个月)。性别,年龄,左主干狭窄,高脂血症,高血压,肾衰竭,目标血管,目标血管狭窄,目标血管直径,手术干预类型,糖尿病,射血分数,术后抗凝方案以及接头尺寸均未显示对旁路移植术通畅性有显着影响(P> .05)。旁路移植物的流动被认为是旁路移植物阻塞的唯一危险因素(P = .0256)。结论:中期随访数据显示大量被闭塞的对称性主动脉连接器系统静脉移植物。基于这些观察,我们机构放弃了使用连接器。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号