首页> 外文期刊>The Journal of oral implantology >Three-dimensional accuracy of implant and abutment level impression techniques: Effect on marginal discrepancy
【24h】

Three-dimensional accuracy of implant and abutment level impression techniques: Effect on marginal discrepancy

机译:种植体和基台水平印模技术的三维精度:对边缘差异的影响

获取原文
获取原文并翻译 | 示例
           

摘要

Impression techniques should precisely represent the 3-dimensional status of implants to allow for the fabrication of passively fitting prostheses and subsequently the elimination of strain on supporting implant components and surrounding bone. The aim of this study was to compare the accuracy of an abutment level impression method with that of an implant level (direct and indirect) impression method using polyether impression material to obtain precise definitive casts and prostheses. A reference acrylic resin dentoform with 2 internal connection implants (Implantium) was made. A total of 21 medium-consistency polyether impressions of the dentoform, including 7 direct implant level, 7 indirect implant level, and 7 abutment level (after 2 straight abutments were secured), were made. Impressions were poured with American Dental Association (ADA) type IV stone, and the positional accuracy of the implant replica heads and abutment analogs in each dimension of x-, y-, and z-axes, as well as angular displacement (Δθ), was evaluated using a coordinate measuring machine. Noble alloy 3-unit castings were fabricated and seated on the abutments in 3 groups; marginal discrepancies were measured at 4 points between prostheses and abutments. Data were analyzed using Mann-Whitney U test, 1-way analysis of variance (ANOVA), and Kruskal- Wallis tests. In comparisons of different impression techniques, only significant statistical Δθ differences were noted between the abutment level method and other techniques (P < .001). Results of this study reveal that although the implant level impression method could better transfer the angular position of the implants (Δθ), the impression method could not affect Δy, Δx, and Δz coordinates of the implants or marginal discrepancy of the 3-unit fixed partial dentures (FPD).
机译:压印技术应准确代表植入物的3维状态,以制造被动装配的假体,并随后消除支撑植入物组件和周围骨骼的应力。这项研究的目的是将基台水平压印方法与使用聚醚压印材料的植入物水平(直接和间接)压印方法的精度进行比较,以获得精确的定型和假体。制备具有两个内部连接植入物(Implantium)的参比丙烯酸树脂牙形。总共制作了21种中等浓度的牙本质聚醚压痕,包括7个直接种植体水平,7个间接种植体水平和7个基台水平(固定2个直基台后)。用美国牙科协会(ADA)的IV型石材浇注印模,并在x,y和z轴的各个维度上植入假体复制头和基台类似物的位置精度,以及角位移(Δθ),使用坐标测量机进行评估。制作了贵金属3单元铸件,并按3组放置在基台上。在假体和基台之间的4个点测量边缘差异。使用Mann-Whitney U检验,方差单向分析(ANOVA)和Kruskal-Wallis检验对数据进行分析。在比较不同的压印技术时,在基台水平方法和其他技术之间仅注意到了显着的统计Δθ差异(P <.001)。这项研究的结果表明,尽管植入物水平压痕方法可以更好地转移植入物的角位置(Δθ),但压痕方法不会影响植入物的Δy,Δx和Δz坐标或固定的3单元固定件的边际差异局部义齿(FPD)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号