首页> 外文期刊>The Journal of laboratory and clinical medicine >Utility of a nitric oxide electrode for monitoring the administration of nitric oxide in biologic systems.
【24h】

Utility of a nitric oxide electrode for monitoring the administration of nitric oxide in biologic systems.

机译:一氧化氮电极用于监测生物系统中一氧化氮的施用的用途。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Amperometric techniques for the detection of nitric oxide (NO) are commercially available, but their sensitivity and specificity are not well described. We evaluated the sensitivity and specificity of a Clark-style, platinum NO electrode. The electrode has a lower limit of detection for NO of <25 pmol/ml in vitro and is linear over the range from 25 pmol/ml to 4 nmol/ml. The electrode is specific for NO so long as the protective membrane that covers the electrode is intact. Any defect in this membrane results in the detection of other redox agents such as hydrogen peroxide. Because of its ease of handling, specificity, and sensitivity, the NO electrode is a useful tool for quantification of administered NO in vitro and in various biologic systems.
机译:用于检测一氧化氮(NO)的安培技术是可商购的,但是其灵敏度和特异性并未得到很好的描述。我们评估了克拉克式铂NO电极的灵敏度和特异性。该电极在体外对NO的检测下限低于25 pmol / ml,并且在25 pmol / ml至4 nmol / ml的范围内呈线性。只要覆盖电极的保护膜完好无损,该电极就专用于NO。该膜上的任何缺陷都会导致检测其他氧化还原剂,例如过氧化氢。由于其易处理,特异性和灵敏性,NO电极是在体外和各种生物系统中定量管理NO的有用工具。