首页> 外文期刊>The Journal of hospital infection >Thermostable adenylate kinase technology: a new process indicator and its use as a validation tool for the reprocessing of surgical instruments.
【24h】

Thermostable adenylate kinase technology: a new process indicator and its use as a validation tool for the reprocessing of surgical instruments.

机译:耐热性腺苷酸激酶技术:一种新的工艺指标,并用作手术器械后处理的验证工具。

获取原文
获取原文并翻译 | 示例
           

摘要

Adenylate kinase (tAK), a thermostable enzyme, was assessed as a possible means of providing a quantitative measure of cleaning efficacy suitable for validating the performance of an automated washer disinfector (AWD) during routine use. Two indicator formulations were developed using either a commercially available washer disinfector soil or a protein-based soil. Each indicator consisted of 100 microg (in test soil) of tAK dried on to a steel or plastic surface. These indicators were placed in each basket of a washer disinfector and processed alongside soiled surgical instruments during a standard day's operation. After processing, remaining tAK activity was detected using a rapid enzyme assay (2 min detection time) in a handheld hygiene monitor. The amount of tAK remaining on each indictor after a full AWD cycle was found to range from 0.1 to 0.4 ng, which represented a mean log(10) removal of 5.8+/-0.3. There was no statistical difference in the residual tAK activity between individual runs or the position of the indicator in the machine. The tAK indicator was also used to analyse the protein removal within each component of the wash cycle. These results demonstrated that all phases of the wash process contributed to the removal of the protein load, with the main wash alone being responsible for 3.6-4.0 log(10) reductions in protein activity. We propose that a quantitative cleaning index using such rapid readout indicator devices would provide a valuable addition to the methodologies for validating cleaning processes.
机译:腺苷酸激酶(tAK)是一种热稳定酶,被评估为一种定量定量的清洁功效的可能手段,适用于验证常规使用中自动洗衣机消毒器(AWD)的性能。使用市售的洗衣机消毒剂污垢或蛋白质基污垢开发了两种指示剂配方。每个指示剂由100微克(在测试土壤中)的tAK组成,并在钢或塑料表面上干燥。将这些指示剂放在洗衣机消毒器的每个篮子中,并在标准的一天操作中与脏污的手术器械一起处理。处理后,使用手持式卫生监测器中的快速酶法检测(2分钟检测时间)检测剩余的tAK活性。在完整的AWD循环后,每个指示器上剩余的tAK量为0.1到0.4 ng,这表示平均log(10)去除为5.8 +/- 0.3。各个运行之间的剩余tAK活性或指示器在机器中的位置之间没有统计学差异。 tAK指示剂还用于分析洗涤循环各组分中的蛋白质去除。这些结果表明,洗涤过程的所有阶段都有助于去除蛋白质负载,其中仅主要洗涤过程就导致蛋白质活性降低3.6-4.0 log(10)。我们建议使用这种快速读数指示器设备的定量清洁指数将为验证清洁过程的方法提供有价值的补充。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号