首页> 外文期刊>The Journal of arthroplasty >Periprosthetic fracture torque for short versus standard cemented hip stems: An experimental in vitro study
【24h】

Periprosthetic fracture torque for short versus standard cemented hip stems: An experimental in vitro study

机译:短髋臼骨与标准髋臼骨干的假体周围骨折扭矩:体外实验研究

获取原文
获取原文并翻译 | 示例
       

摘要

In an attempt to preserve proximal femoral bone stock and achieve a better fit in smaller femora, especially in the Asian population, several new shorter stem designs have become available. We investigated the torque to periprosthetic femoral fracture of the Exeter short stem compared with the conventional length Exeter stem in a Sawbone model. Forty-two stems; 21 shorter and 21 conventional stems both with three different offsets were cemented in a composite Sawbone model and torqued to fracture. Results showed that Sawbone femurs break at a statistically significantly lower torque to failure with a shorter compared to conventional-length Exeter stem of the same offset. Both standard and short-stem designs are safe to use as the torque to failure is 7-10 times that seen in activities of daily living.
机译:为了保护股骨近端骨股并更好地适合较小的股骨,尤其是亚洲人群,已有几种新的较短的股骨设计可供选择。我们比较了在Sawbone模型中与传统长度的Exeter茎相比,Exeter短茎的假体周围股骨骨折的扭矩。四十二茎;在复合的Sawbone模型中,将21个较短的茎和21个常规茎(均具有三个不同的偏移量)固结在一起,并拧紧以使其断裂。结果显示,锯齿状股骨的断裂扭矩在统计学上显着降低,与传统长度的相同偏移量的埃克塞特杆相比,断裂力更短。标准和短杆设计均可安全使用,因为破坏扭矩是日常生活活动的7至10倍。