首页> 外文期刊>The Journal of arthroplasty >Computerized range of motion analysis following dual mobility total hip arthroplasty, traditional total hip arthroplasty, and hip resurfacing
【24h】

Computerized range of motion analysis following dual mobility total hip arthroplasty, traditional total hip arthroplasty, and hip resurfacing

机译:双活动全髋关节置换术,传统全髋关节置换术和髋关节置换术后的运动分析的计算机范围

获取原文
获取原文并翻译 | 示例
           

摘要

Newer arthroplasty designs claim to provide superior range of motion (ROM) and greater stability than their predecessors. However, there is no way to compare ROM of implant systems in an equivalent anatomical environment in a clinical setting. This study used computer-aided design to compare ROM after hip resurfacing, 28. mm THA, 36. mm THA, and anatomic dual mobility (ADM) THA in 3D models of 5 cadaver pelvises. ROM to impingement was then tested in 10 different motions and a one-way ANOVA was used to compare results. The hip resurfacing resulted in restricted ROM compared to the other 3 models in all motions except adduction. The ADM, 36. mm, and 28. mm THA resulted in similar ROM. Dual mobility constructs provide comparable ROM in patients where large head THA is not appropriate.
机译:较新的关节成形术设计声称,与以前的产品相比,可提供出色的运动范围(ROM)和更高的稳定性。但是,在临床环境中,没有办法在等效的解剖环境中比较植入系统的ROM。这项研究使用计算机辅助设计比较了5具尸体骨盆的3D模型中髋关节表面置换,28。mm THA,36。mm THA和解剖学上的双活动性(ADM)THA后的ROM。然后以10种不同的运动测试ROM的碰撞情况,并使用单向方差分析比较结果。与其他3个模型相比,髋关节表面置换术在除内收外的所有动作中均限制了ROM。 THA为36.mm和28.mm的ADM产生了相似的ROM。在不适合使用大头颅THA的患者中,双重移动性构造可提供可比的ROM。