Effect of femoral component design on patellofemoral crepitance and patella clunk syndrome after posterior-stabilized total knee arthroplasty

机译:股骨组件设计对后路稳定全膝关节置换术后of股肌pit发和骨陈旧综合征的影响

获取原文
获取原文并翻译 | 示例
           

摘要

The purpose of this study was to determine if recent changes to the femoral component of a particular posterior-stabilized total knee prosthesis would affect the incidence of postoperative patellofemoral crepitance and patella clunk syndrome. One hundred eight total knee arthroplasties were performed with the conventional design; 136 were performed after the femoral component was changed. Complications were compared between the groups with an average follow-up of 17.7 months and 12.4 months, respectively. Thirteen knees with the conventional design (12%) were found to have patellofemoral complications; no complications were noted with the new design (P < .0001). Femoral components with a deep trochlear groove and smooth transition of the intercondylar box appear to better accommodate any peripatellar fibrous nodule that may form after total knee arthroplasty.
机译:这项研究的目的是确定特定后路稳定全膝关节假体的股骨组件最近发生的变化是否会影响术后pa股厚和骨旧骨综合征的发生率。采用常规设计进行了一百零八次膝关节置换。股骨组件更换后进行了136例。比较两组之间的并发症,平均随访时间分别为17.7个月和12.4个月。常规设计的十三个膝盖(12%)被发现有pa股并发症。新设计未发现并发症(P <.0001)。股骨组件具有较粗的滑车槽和con间盒平滑过渡,似乎可以更好地适应全膝关节置换术后可能形成的pat骨周围纤维结节。