首页> 外文期刊>The Journal of arthroplasty >Migration of the trabecular metal monoblock acetabular cup system.
【24h】

Migration of the trabecular metal monoblock acetabular cup system.

机译:小梁金属整体式髋臼杯系统的迁移。

获取原文
获取原文并翻译 | 示例
           

摘要

Fifty-one primary total hip arthroplasties were performed using Trabecular Metal Monoblock Acetabular Cup System (Zimmer Inc, Warsaw, IN). In a 2-year prospective study, its behavior was closely monitored by clinical, radiologic, and component migration assessment by Ein-Bild-Rontgen-Analyse method. There were no complications. The mean Harris hip score was improved from 41 to 95. A polar gap at the postoperative radiograph was noted at 25% of the arthroplasties, most of which subsequently filled with bone within 6 months. The mean total absolute migration was 0.67 mm. The implant showed excellent early clinical and radiographic behavior. The 2-year migration rate study, as index of long-term survival and success, showed very good early implant stability and, in all cases except one, superior to the available studies for similar design acetabular cups.
机译:使用小梁金属整体式髋臼杯系统(Zimmer Inc,华沙,印第安纳州)进行了五十一次主要的全髋关节置换术。在一项为期2年的前瞻性研究中,通过Ein-Bild-Rontgen-Analyse方法通过临床,放射学和组分迁移评估对其行为进行了密切监测。没有并发症。 Harris髋关节的平均得分从41提高到95。术后X射线照片显示极间隙出现了25%的关节置换术,其中大部分在6个月内充满了骨头。平均总绝对偏移为0.67毫米。植入物显示出优异的早期临床和放射学行为。这项为期两年的迁移率研究作为长期存活和成功的指标,显示出非常好的早期植入物稳定性,并且在所有情况下(除一种情况下)都优于类似设计的髋臼杯的现有研究。