Oxinium exhibits 4900 times less volumetric wear and 640 times less deep scratches in laboratory wear testing compared with traditional cobalt-chromium alloys. Despite the superior wear resistance because of the thin ceramic surface, the zirconium alloy substrate is relatively soft (Hv = 285) when compared with cobalt-chrome alloy femoral heads (ISO 5832-12; Hv = ~420,) and may deform in contact with acetabular shell materials (ISO5832-3; Hv = 350) in the case of dislocation. Three retrieval cases highlight the damage that may occur during dislocation and/or when performing closed reduction maneuvers. It is recommended that closed reduction be attempted with caution because significant head damage can occur. This may lead to accelerated polyethylene wear. In the case of successful reduction, close patient follow-up is recommended. Open reduction with femoral head inspection and exchange may be preferable if difficulty is encountered in closed reduction maneuvers.
机译:与传统的钴铬合金相比,在实验室磨损测试中,氧的体积磨损小4900倍,深划痕少640倍。尽管由于陶瓷表面薄而具有优异的耐磨性,但与钴铬合金股骨头(ISO 5832-12; Hv =〜420,)相比,锆合金基材相对较软(Hv = 285),并且可能在接触时变形脱臼时使用髋臼壳材料(ISO5832-3; Hv = 350)。三种取回案例突出了在脱位和/或执行封闭复位操作时可能发生的损坏。建议谨慎进行闭合复位,因为这可能会严重损坏头部。这可能会导致聚乙烯磨损加速。如果成功减少,建议密切随访患者。如果在封闭复位手术中遇到困难,则最好进行股骨头检查和置换的开放复位。