首页> 外文期刊>The Journal of arthroplasty >The porous-coated anatomic total hip prosthesis: 11- to 13-year results.
【24h】

The porous-coated anatomic total hip prosthesis: 11- to 13-year results.

机译:多孔涂层解剖型全髋关节假体:11至13年的结果。

获取原文
获取原文并翻译 | 示例
           

摘要

Of 137 consecutive primary porous-coated anatomic (PCA) arthroplasties performed, complete data was available on 107 hips in 93 patients at an average of 12.4 years (range, 11-13 years). The average Harris Hip Score improved from 44 to 85 at final follow-up, with 78% good or excellent result. Thirteen percent (14 of 107) of acetabular components and 4 porous ingrowth stems were deemed failures. Osteolytic lesions were seen in 7.5% (8 of 107) of acetabular components. Survivorship of the PCA prosthesis as a whole was 83% at 12.4 years. This first-generation femoral stem performed well, with only 4 failures, all of which occurred in patients with avascular necrosis in which unrecognized pathologic changes in the proximal femur may have inhibited bone ingrowth. The majority of the failures can be attributed to aseptic loosening of the acetabular component.
机译:在进行的137次连续的原发性多孔涂层解剖(PCA)关节置换术中,有93位患者的107髋的完整数据可用,平均为12.4年(范围11-13岁)。在最后的随访中,Harris Hip的平均得分从44分提高到85分,有78%的好或好结果。髋臼组件中有13%(107个中的14个)和4个多孔的向内生长茎被认为是失败的。在7.5%(107个中的8个)髋臼组件中可见溶骨性病变。整个PCA假体在12.4年时的存活率为83%。第一代股骨柄表现良好,仅有4例失败,所有这些都发生在无血管坏死的患者中,其中未识别的股骨近端病理改变可能抑制了骨向内生长。大多数失败可归因于髋臼组件的无菌性松动。